Open The Measuring Instrument - TESTO Next-Gen 300 Instruction Manual

Combustion analyzer
Hide thumbs Also See for Next-Gen 300:
Table of Contents

Advertisement

10 Maintenance
1
Open the condensate outlet on the
condensate container.
2
Let the condensate run out into a
sink
3
- Wipe off any drops still on the condensate outlet with a cloth and close
the condensate outlet.
The condensate outlet must be completely closed, otherwise incorrect
measurements could occur due to infiltration of external air.

10.6 Open the measuring instrument

Only open the measuring instrument when this is required for maintenance
purposes (replacing gas sensors).
The measuring instrument must not be connected to a mains socket via
the mains unit. The measuring instrument must be switched off.
When opening/assembling the instrument, take care not to lose any
screws that have been removed. It is advisable to place a cloth on the
work surface.
1
Place the instrument on its front so that the back is facing upwards.
83

Advertisement

Table of Contents
loading

Table of Contents