11.Connect one end of a wire clip lead to a small screwdriver having a
12.Using a small screwdriver with the shank in contact with the CRT
13.Pry out the black grommet protecting post-accelerator cable
14.Carefully unsnap the A6A1W3 post-accelerator cable from the CRT
15.Place the spectrum analyzer on its right side frame with the front
16.Fold out the A14 and A15 assemblies as described in steps 3 and 4
The voltage potential at A6A1W3 is +9 kV. Failure to discharge
WARNING
A6A1W3 correctly may result in severe electrical shock to
personnel and damage to the instrument.
Chapter 3
conductive shank and tip. Connect the other end of the clip lead to
the CRT shield assembly as shown in
the insulated screwdriver handle and slip the tip of the screwdriver
under the rubber shroud of the A6A1W3 post-accelerator cable,
shorting the cable to ground through the CRT shield assembly. See
Figure 3-3 on page
162.
shield assembly, slip the tip of the screwdriver under the A6A1W3
post-accelerator cable rubber shroud and short the cable to ground
on the CRT shield assembly.
A6A1W3 from the CRT shield assembly.
and discharge it by shorting the cable to chassis ground on the CRT
shield assembly.
frame assembly hanging over the front edge of the workbench.
under "Procedure 9. A14 and A15 Assemblies Removal."
Assembly Replacement
Procedure 2B. A1 Front Frame/A18 CRT (8560E)
Figure 3-3 on page
162. Hold
161