Procedure

1. Press
analyzer cover and place the spectrum analyzer in the service
position as illustrated in
Figure 2-14
Tracking Generator Power Setup and Adjustment Locations
2. Connect the Type N cable between the RF OUT 50Ω and RF INPUT
50Ω connectors on the HP 8560E/EC.
3. Press
Chapter 2
9. Tracking Generator Power Level Adjustments (Option 002)
to turn the spectrum analyzer off. Remove the spectrum
LINE
on the HP 8560E/EC and set the controls as follows:
PRESET
Center frequency ............................................. 300 MHz
Span..................................................................... 0 Hz
Figure 2-14 on page
Adjustment Procedures
95.
95