Procedure

1. Remove the A10 tracking generator assembly as described in
Chapter 3. With the A10 tracking generator positioned next to the
HP 8560E/EC, reconnect W14 (10-wire ribbon cable) to A10J1.
Reconnect W48 to A10J8. Connect the 50 Ω termination to A10J3.
2. Connect the equipment as shown in
frequency counter provides the frequency reference for the
synthesized sweeper and the HP 8560E/EC.
Figure 2-25
Tracking Oscillator Adjustment Setup
Chapter 2
Adjustment Procedures
18. Tracking Oscillator Adjustment (Option 002)
Figure 2-25 on page
123. The
123