Procedure
1. Press
to turn the spectrum analyzer off and disconnect the ac
LINE
power cord. Remove the spectrum analyzer cover and connect the
equipment as illustrated in
power cord and press
2. Set the HP 3456A controls as follows:
Function .................................................................... DC VOLTS
Range .................................................................................. 0.1 V
Resolution ........... .......................................................... 100 mV
3. On the HP 8560E/EC, press
.
BIAS OFF
4. Adjust A15R926 EXT BIAS ZERO for a DVM reading of 0.000 Vdc
±12.5 mV.
Chapter 2
14. External Mixer Bias Adjustment
Figure 2-20 on page
to turn the spectrum analyzer on.
LINE
,
AUX CTRL
EXTERNAL MIXER
Adjustment Procedures
108. Reconnect the
,
, then
BIAS
109