HP 8560E Service Manual page 36

Spectrum analyzers
Table of Contents

Advertisement

Table 1-3
Recommended Test Equipment
Instrument
Critical Specifications for Equipment Substitution
Counters
Frequency
Output frequency: 10 MHz
standard
Accuracy: <1 × 10
Microwave
Frequency range: 9 MHz to 26.5 GHz
frequency
Timebase accuracy (aging): <5 × 10
counter
External frequency reference input
Universal
Modes: TI A→B, frequency count
counter
Time interval measurement range: 100 ns to 120 s
Frequency count range: 400 Hz to 11 MHz
Frequency resolution: 1 mHz
Timebase accuracy (aging): <3 × 10
External 10 MHz reference input
Receivers
Spectrum
Frequency range: 300 kHz to 7 GHz
analyzer
(for HP
Relative amplitude accuracy:
8560E/EC
Option 002)
300 kHz to 2.7 GHz: <±1.8 dB
300 kHz to 7 GHz: <±4.0 dB
Absolute amplitude accuracy:
3.9 GHz to 6.9 GHz: <±2.7 dB
Frequency accuracy:
<±10 kHz at 7 GHz
Spectrum
Frequency range: 300 kHz to 7 GHz
analyzer
Amplitude range: −70 dBm to +20 dBm
Measuring
Compatible w/power sensors
receiver
dB relative mode
Resolution: 0.01 dB
Reference accuracy: <±1.2%
* Part of microwave workstation
P = performance tests; A = adjustments; M = test & adjustment module;
T = troubleshooting;
V = operation verification
Chapter 1
−10
−10
/day
−7
/month
General Information
Recommended Test Equipment
Recommended
Use
Model
P,A
HP 5061B
P,A,M,V
HP 5343A*
Option 001
HP 5334A/B
P
HP 8566B*
P,A,T
A,T
HP 8566B*
HP 8902A*
P,A,T,
M,V
35

Advertisement

Table of Contents
loading

This manual is also suitable for:

8560ec

Table of Contents