General Information
Recommended Test Equipment
Table 1-3
Recommended Test Equipment
Instrument
Critical Specifications for Equipment Substitution
Sensors
Power sensor
Frequency range: 10 MHz to 12 GHz
Maximum SWR:
1.40 (10 to 30 MHz)
1.18 (30 to 50 MHz)
1.10 (50 MHz to 2 GHz)
1.18 (2 to 12.4 GHz)
Power sensor
Frequency range: 250 MHz to 350 MHz
Power range: 100 nW to 10 µW
Maximum SWR: 1.15 (250 to 350 MHz)
Power sensor
Frequency range: 100 kHz to 2.9 GHz
Maximum SWR:
1.1 (1 MHz to 2.0 GHz)
1.30 (2.0 GHz to 2.9 GHz)
Other
Equipment
Controller
Required to run operation verification software
and adjustment/diagnostic software (HP 8564E/
HP 8565E)
Oscilloscope
Bandwidth (3 dB): dc to 100 MHz
Two channels
Minimum vertical deflection factor: ≤5 mV/div
Minimum timebase setting: <100 ns
Digitizing display with time cursors
Delta-t cursor accuracy in 500 ns/Div: <0.1 µs
Amplifier
Frequency range:
HP 8560E/EC, 2.0 to 2.9 GHz
HP 8561E, 2.0 to 6.5 GHz
HP 8563E, 2.0 to 8.0 GHz
HP 8564E, 2.0 to 8.0 GHz
HP 8565E, 2.0 to 8.0 GHz
Minimum output power (leveled)
2.0 to 8.0 GHz: +16 dBm
* Part of microwave workstation
P = performance tests; A = adjustments; M = test & adjustment module; T = troubleshooting;
V = operation verification
36
Recommended
Use
Model
P,A,T,
HP 8481A*
M,V
HP 8481D
P,A
HP 8482A*
P,A,T,
M,V
HP 9816A,
V
HP 9836A/C,
HP 310, 320
HP 332, 360
HP 54501A*
P,A,T
P
HP 11975A
Chapter 1