Installation; Ports Required For Communication; Installation Steps - LINET LINIS SafetyPort Instructions For Use Manual

System for automated transfer and visualisation of various data from linet beds
Table of Contents

Advertisement

7

Installation

The LINIS SafetyPort system is installed by LINET. The standard installation is performed using remote access, that is
provided by the customer (VPN). VPN provision is a required and installation cannot be completed unless it is provided.
VPN access shall be handled by a separate NDA (provided by a customer).

7.1 Ports required for communication

 bed
LINIS SafetyPort server: 10180, LINIS SafetyPort server
←→
 user workstation
 LINIS SafetyPort server

7.2 Installation steps

Installation prerequisites:
Functional internet infrastructure
Functional hospital information system
Opened required ports
Stable display monitor to open Administration interface and/or SafetyPort Dashboard
Ready to install LINIS server
Remote access to the server
Installation of beds
Installation steps:
LINIS server installation
Communication of Multicare beds or Multicare LE beds or Eleganza 5 beds or Eleganza 4 or Eleganza 3XC
or Eleganza 3 (US version) beds with LINIS server
Localisation system installation (LINIS SafetyPort bed + Tag)
SafetyPort installation
Test of the functional connection of bed, server and 3rd party software (HIS)
Technical training
20/52
LINIS SafetyPort server: 8888, 3600, 3306
←→
third party server (e.g: HIS): based on third party server requirements (1337 typically)
bed: 10180 or any
D9U001SW2-0110_03