Mechanical installation
7
Electrical connection
14
Pin assignment
1
2
3
4
5
6
7
If the transducer is connected in accordance with the fol
lowing connection diagram, the output voltage at the
amplifier is positive when under compressive loading.
CAUTION
Caution with very long cables.
Cable connection
Only use shielded and low-capacitance measurement
cables. Ensure that the connection is perfect and that it
has a low contact resistance. We recommend implement
ing extensions using six-wire configuration to exclude
changes to the sensitivity.
A4303-1.0 HBM: public
Device plug
Measurement signal
(+)
Excitation voltage (-)
Excitation voltage (+) bu (blue)
Measurement signal
(-)
Shield
Sensor circuit (+) to
3
Sensor circuit (-) to 2
Cable assignment
wh (white)
bk (black)
rd (red)
gn (green)
gy (gray)
K-LK1/600N-WH