4. Measuring Thickness
Step 3: Press
below the desired column to move the
cursor from one character to another on the VIRTUAL KEY-
BOARD. When the desired character is selected, press
to send the letter to the input line. When you've finished
imputing the parameter set name, press
P-SET naming process. Note that your newly-saved parameter
set is now listed when you scroll through the list of parameter
sets by pressing
under the
P-SET selection. Note also that specifying a parameter set
name that's the same as an existing set will NOT cause the
existing set to be overwritten. The existing set must first be
deleted.
Loading a Parameter Set
When you're performing an application for which you've
already saved a parameter set, you can apply that parameter
set by using the PARAM Submenu. This is accessed from the
TG Primary Menu (Figure 3-4).
Page 132
Step 1: Activate the PARAM Submenu (located in the TG
Primary Menu) by selecting it with
highlighted. Five selections will appear across the bottom of the
display screen.
Step 2: Press
to complete the
display a list of all the saved parameter sets.
Step 3: When the desired parameter set is displayed under the
P-SET selection, press
P-SET. This will cause the saved parameter set to be applied
to the instrument, which will automatically adjust all of the
features controlled by a parameter set to the setting defined in
the selected P-SET. (Refer to Table 4-1.)
NOTE: Before continuing, activate the TGMODE Submenu
and confirm that the probe type matches the one connected
to the instrument.
or
below the selection titled P-SET. This will
below the selection titled LOAD
DMS 2 Operating Manual
so that it's