10 - 50
1.
Press Start/Hold on the injector head, the workstation, or the handswitch.
NOTE: Refer to Table 6 - 1 for description on how the Arm Lights function while armed, injecting, and/or
during a hold.
NOTE: If a Reminder is set, a communication appears once the Reminder elapsed time equals the setting.
If a test inject phase is programmed, the Reminder elapsed time begins with the start of the first
phase after the test inject.
NOTE: The Reminder communication is removed when either the communication is acknowledged or
another communication is activated.
NOTE: Test Injection - an injection phase followed by a hold phase. After the injection phase, the system
holds the injection until Start/Hold is pressed on the injector head, workstation, or handswitch to
resume the protocol.
NOTE: Hold Phase - If the protocol contains a Hold Phase, the system holds the injection until Start/
Hold is pressed on the injector head, workstation, or handswitch to resume the protocol.
10.4 Operator-Initiated Hold
If Start/Hold is pressed during an injection phase, the system holds the injection until Start/Hold is pressed on the
injector head, workstation, or handswitch to resume the protocol.
NOTE: If a hold is initiated, the Reminder elapsed time is halted. The Reminder elapsed time starts when
the protocol is reinitiated.
NOTE: If the system remains in a hold condition for (20) minutes, the system aborts the injection
automatically.
10.5 Injection Aborted
At any time, press the Abort button or any other button on the injector head (other than Start/Hold) to abort the
injection.
10.6 Viewing Injection Progress
During an injection, the system displays the following:
NOTE: If the injection is aborted, all remaining reminders are ignored.
MEDRAD® Stellant FLEX CT Injection System with Certegra® Workstation