MEDRAD
Stellant CT Injection System with Certegra
NOTE: Hold Phase - If the protocol contains a Hold Phase, the system holds the injection until the
10.4 Operator-initiated Hold
If the operator presses Start/Hold during an injection phase, the system holds the injection until the
operator presses Start/Hold on the injector head, workstation, or handswitch to resume the protocol.
NOTE: If an operator initiates a hold, the Reminder elapsed time is halted. The Reminder elapsed
NOTE: If the system remains in a hold condition for (20) minutes, the system aborts the injection
10.5 Injection Aborted
At any time, press the Abort button or any other button on the injector head (other than Start/Hold) to
abort the injection.
10.6 Viewing Injection Progress
During an injection, the system displays the following:
NOTE: If the injection is aborted, all remaining reminders are ignored.
1
2
3
4
5
10 - 44
operator presses Start/Hold on the injector head, workstation, or handswitch to resume the
protocol.
time starts when the operator re-initiates the protocol.
automatically.
Figure 10 - 4: Injection Views
The graph shows the pressure sensed by the system during an
Pressure Graph
injection.
Phase View
The system highlights each phase as it starts.
Shows the duration of the injection.
Elapsed Time
•
Pressure Limit
Shows the programmed pressure limit.
Reminders
Indicates the number of set reminders.
®
Workstation
If a test injection is programmed, the elapsed time begins
after the test injection.