The following is a list of error messages the injection system could display and corrections for those issues.
Error Message
Related Warning
Expired syringe
detected on side <A or
B>.
WARNING: Syringe reuse
may result in biological
Previously installed
contamination, product
syringe detected on
degradation, and/or
side <A or B>.
product performance
issues.
Syringe interface error
on side <A or B>.
MEDRAD® Stellant FLEX CT Injection System with Certegra® Workstation
Table 14 - 1: Error Screen Messages
Explanation
The system compares the syringe manufacturing date to
the date the syringe is installed on the system. When a
syringe is past the expiration date, an error message
informs the user. If an expired syringe is installed and the
error message occurs, then:
Confirm that the system's date and time are cor-
rect. Refer to
instructions.
Remove the existing syringe and install a new
syringe that is not expired, or
Press OK to override the message and use the
current syringe.
The system tracks the number of times a syringe has been
installed. If a syringe has been installed and removed
multiple times, an error message occurs. To correct the
error:
Confirm that the syringe is being installed cor-
rectly. Refer to
instructions on how to correctly install a syringe,
or
Remove the existing syringe and install a new
syringe. Properly discard disposable items after
use, or if there is any possibility that contamina-
tion may have occurred, or
Press OK to override the message and use the
current syringe.
An error message can occur for any of the following
reasons:
A defective syringe is used,
A syringe is incorrectly inserted on the system, or
The injector is not correctly identifying the
syringe.
To correct the error:
Remove the existing syringe and install a new
syringe. Refer to
instructions on how to install a syringe.
A software update may be required. Contact
Bayer.
If installing a new syringe from Bayer does not correct the
problem, record the error code and contact Bayer.
13.1 System Setup
for further
9.2.2 Install and Fill a Syringe
9.2.2 Install and Fill a Syringe