18 - 102
Guidance and manufacturer's declaration - electromagnetic immunity
3Vrms from 80 MHz to 2.7 GHz at 80% AM 1kHz and
specific ISM bands listed below:
Frequency
(MHz)
385
450
710
745
Radiated RF
780
IEC 61000-4-3
810
870
930
1720
1845
1970
2450
5240
5500
5785
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
system is used exceeds the applicable RF compliance level above, the system should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the system.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
MEDRAD® Stellant FLEX CT Injection System with Certegra® Workstation
Field
Modulation
Modulation
Strength
Type
Frequency
(Volts/
meter)
Pulse
18 Hz
27
Pulse
18 Hz
28
Pulse
217 Hz
9
Pulse
217 Hz
9
Pulse
217 Hz
9
Pulse
18 Hz
28
Pulse
18 Hz
28
Pulse
18 Hz
28
Pulse
217 Hz
28
Pulse
217 Hz
28
Pulse
217 Hz
28
Pulse
217 Hz
28
Pulse
217 Hz
9
Pulse
217 Hz
9
Pulse
217 Hz
9
d
=
1.17 p
80 MHz to 800 MHz
d
=
2.33 p
800 MHz to 2.7 GHz
Where p is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in each
frequency range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Non-ionizing Radiation Symbol (IEC TR
60878, 5140)