Fluid Delivery Setup Configurable Items - Bayer HealthCare CERTEGRA MEDRAD Stellant FLEX Operation Manual

Ct injection system
Table of Contents

Advertisement

Operation Manual

13.4.1 Fluid Delivery Setup Configurable Items

Configurable Item
Description
Sets the injector to prime both syringes simultaneously or one after the other.
The Simultaneous mode controls fluid movement through the union connector,
moving both saline and contrast simultaneously during the priming process.
Priming Mode
The Sequential mode controls fluid movement through each branch of the union
connector, moving saline and contrast separately, one after the other during the
priming process.
Sets the syringe to be used for priming the tubing. Both syringes are primed. For a test
Priming Source
inject, the operator may want to select the saline syringe as the priming source.
Sets the type of Low Pressure Connector tubing being used.
LPCT Type
Test Inject Source
Sets the default syringe source for the test injection.
Test Inject Volume
Sets the default test injection volume.
Test Inject Rate
Sets the default flow rate for the test injection.
The system calculates values for the displayed parameter. Sets the Programming
Mode to enable the system to calculate the duration, flow rate, or volume.
Protocol Programming
The Programming Modes follow:
Mode
Sets the total volume to display the volumes in Syringe A and Syringe B, or the total in
Total Volume Display
Syringe A and B combined.
Enables or disables the Auto Retract feature to retract the piston after a syringe is
Auto Retract
removed.
Enable or disables the Auto Advance feature to advance the piston after a syringe is
Auto Advance
installed.
When using filling devices from Bayer, choose one of the four volume selections based
on the following guidelines:
Auto Load Purge
Volume
For other filling devices, consult manufacturer's instructions for use.
Fluid Control
Sets the resistance level on the manual knobs.
Contrast Types
Set or add types of contrast.
Table 13 - 3: Configurable Items
NOTE: Selecting the Other LPCT configuration setting will provide about 6
mL of prime fluid when using one syringe, or about 6.5 mL when
using two syringes.
Calculate Flow Rate: SYSTEM CALCULATES THE FLOW RATE based on volume and duration.
Calculate Volume: SYSTEM CALCULATES THE VOLUME based on flow rate and duration.
Calculate Duration: SYSTEM CALCULATES THE DURATION based on flow rate and volume.
30 mL for QFT
20 mL for short transfer sets (23" / 58 cm)
10 mL when loading through a spike only
0 mL (OFF) when loading through a filling device that has a check valve