Bayer HealthCare CERTEGRA MEDRAD Stellant FLEX Operation Manual page 45

Ct injection system
Table of Contents

Advertisement

Operation Manual
DualFlow automatically determines the individual volumes of Contrast and Saline and the individual flow
rates needed for the Contrast and Saline portions of the DualFlow phase.
For example, if the user enters a total Flow Rate of 4.0 mL/sec and a total Volume of 60 mL, the individual
volume and flow rate of Contrast and Saline for the DualFlow phase would be as follows (based on the
sample percentages of Contrast):
Sample Ratio
A90% B10%
A80% B20%
A75% B25%
A70% B30%
A65% B35%
A60% B40%
A50% B50%
NOTE: To change the ratio once the phase has been programmed, re-select the phase and enter the new
ratio values per step 3.
Syringe A
54 mL @ 3.6 mL/sec
48 mL @ 3.2 mL/sec
45 mL @ 3.0 mL/sec
42 mL @ 2.8 mL/sec
39 mL @ 2.6 mL/sec
36 mL @ 2.4 mL/sec
30 mL @ 2.0 mL/sec
Syringe B
6 mL @ 0.4 mL/sec
12 mL @ 0.8 mL/sec
15 mL @ 1.0 mL/sec
18 mL @ 1.2 mL/sec
21 mL @ 1.4 mL/sec
24 mL @ 1.6 mL/sec
30 mL @ 2.0 mL/sec
8 - 37