Initiating An Injection; Operator Initiated Hold; Aborting An Injection - Bayer HealthCare Medrad MRXperion Operating Manual

Mr injection system
Hide thumbs Also See for Medrad MRXperion:
Table of Contents

Advertisement

12 - 52
1.
Following the I Checked for Air Confirmation, select the KVO button to initiate KVO.
2.
KVO may be stopped at any time by selecting the KVO button after KVO is initiated (illuminated in blue) or by
pressing any injector head control button. This will also disarm the system and terminate any injection in progress.
NOTE: Other actions that disarm the injector, such as syringe removal, pressing the Disarm button, and injection
stall will also stop KVO.
NOTE: If an occlusion occurs during KVO, the system will detect the condition after 4 or less KVO boluses fail to be
delivered. (The occlusion will be detected after 1 minute or less if the KVO interval is set to 15 seconds or
after 5 minutes or less if the KVO interval is set to 75 seconds.) Refer to
for information about setting KVO intervals.

12.4 Initiating an Injection

NOTE: It is recommended that the operator stay by the patient's side at the beginning of the injection.
1.
Press Start/Hold on the injector head, Pod, or hand switch.
NOTE: See
"Chapter 7 - System Lights and Indicators"
armed, injecting, and/or during a hold.
NOTE: If the protocol contains a Hold phase, the system holds the injection until the operator presses Start/Hold
on the injector head, Pod, or hand switch to resume the protocol.

12.5 Operator Initiated Hold

If the operator presses Start/Hold during an injection phase, the system holds the injection until the operator presses Start/
Hold on the injector head, Pod, or hand switch to resume the protocol.
NOTE: If an operator initiates a hold, the Reminder elapsed time is halted. The Reminder elapsed time starts when
the operator re-initiates the protocol.

12.6 Aborting an Injection

At any time, press the Abort button on the Pod or injector head to abort the injection. An injection will also abort if the
injector head is moved to the upright position during an injection.
"Section 15.4 - Fluid Delivery Setup"
for a description of how the system lights function while
MEDRAD® MRXperion