Removing Disposables

WARNING
Biological Contamination Hazard - Serious patient and/or worker injury or death may result.
Properly discard fluid source containers and disposable items after use (refer to disposable label for
specifics), or if there is any possibility that contamination may have occurred.
Do not reuse single-use fluid sources. Refer to the fluid source manufacturers' instructions for use.
1.
Disconnect the disposable tubing set from the catheter.
NOTE: The disposable tubing set does not need to be disconnected from the syringe.
2.
Remove the syringe by twisting counterclockwise ¼ turn and pulling out.
NOTE: If the syringe cannot be removed, rotate the injector head manual knob approximately two
revolutions in the reverse direction, then repeat step 2 above.
NOTE: Once syringe has been removed, the piston automatically retracts to the start position (if
enabled).
3.
Discard fluid source containers and all used disposable items in accordance with facility policy.