Material:
Clean, soft, non-linting wipe
Warm water
Cleaning Agent:
Use either Sani-Cloth® Plus Germicidal Disposable Cloth (EPA registration # 9480-6) or CaviWipes™
Disinfecting Towelettes (EPA registration # 46781-8) within the United States.
Use either Sani-Cloth® Plus Germicidal Disposable Cloth or CaviWipes™ Disinfecting Towelettes in all
other countries. If neither Sani-Cloth® Plus Germicidal Disposable Cloth nor CaviWipes™ Disinfecting
Towelettes are available within your country then use a disinfectant with 0.25% - 0.30% quaternary
ammonium compounds and 14.8% - 20% isopropanol.
1.
Wipe the workstation screen with a clean, soft, non-linting wipe dampened with warm water for 1 minute.
2.
Wipe the workstation screen with Sani-Cloth® Plus or CaviWipes™ for 1 minute or until visibly clean.
NOTE: If in a country where Sani-Cloth® Plus or CaviWipes™ are not available, and using an equivalent
cleaning agent (refer to the materials listed above), follow the same instructions specified in
step 2.
3.
Inspect the workstation screen to ensure all surfaces are clean.
4.
If any visible contrast media or other debris is evident, repeat the cleaning instructions steps 1 through 3
until no visible soil is detected.

15.4 Monthly

Once a month, perform an operational checkout.

15.4.1 Operational Checkout

A basic functional checkout of the system should be included as part of regular maintenance. Verifying proper
operation of the system will help in detection of any problems that may not be noticed in day-to-day operation. The
following procedure represents a suggested series of activities which encompass typical operation of the system.
Read the following procedure carefully before beginning the checkout. If problems are detected, call your Services
Representative.
NOTE: Any problems detected during this or any other procedure should be addressed before using the
system in patient procedures.
System Labels
Ensure that all system safety and warning labels are in place and legible.
Power Up
1.
Apply power to the system.
2.
Verify the Safety screen displays.
3.
Press CONTINUE to acknowledge the messages on the Safety screen.
4.
Verify proper audible tones are functioning on the workstation and injector head.
5.
Verify all displays and indicators are functioning properly.
6.
Verify the arm lights on the injector head are functioning.
Programming
After the Main screen displays, verify that the following controls are functioning properly.
15 - 83