Stratasys J826
Environmental Conditions
Temperature and Humidity
The temperature and relative humidity around the printer must be maintained within certain limits. Peak
conditions occur when the printer and the server workstation draw maximum electrical power, resulting in
the heat dissipation listed in the table below.
Printer
Server workstation
Total
The room temperature and relative humidity should not exceed the limits listed in the table below.
Air Quality
To ensure optimal air quality around the printer, a connection to an external ventilation duct is
recommended (see
should change the air at least four times every hour.
Noise Level
The noise level around the printer is typically less than65 dB during printing.
©
DOC-18005 Rev. E
Table 5 Heat dissipation (peak conditions)
Table 6 Room temperature and relative humidity specifications
Temperature
Relative humidity
"Accessories and Utilities" on the next
2019-2020 Stratasys. All rights reserved.
Heat Dissipation
1500 W (5140 BTU/hr)
450 W (1530 BTU/hr)
1950 W (6670 BTU/hr)
Range
18°C to 25°C (64.5°F to 77°F)
30%–70% non-condensing
page). Otherwise, the room ventilation system
Environmental Conditions
16