Specifications
9.1

Performance Specifications

Type
Temperature range
Temperature accuracy
Spatial temperature deviation
Plate capacity
Sample capacity
Cycle time for handling system
PC connection
Control
Safety features
Good Laboratory Practice (GLP)
features
Page 74
The Charger performance is specified as follows:
Specification
4-40 °C (settable),
cooling target ≥22 K below ambient
-2 °C / +4 °C
±2 °C
9 deep well plates, maximum height 45 mm or
9 sample racks, maximum height 45 mm including vials or
20 well plates, maximum height 20 mm
Up to 7,680 samples in twenty 384 well plates
< 1 min (shuttling a sample rack or well plate from or to autosampler)
USB 2.0 interface
Chromeleon 7, Chromeleon 6.8
The Charger can be operated also with other data systems. For details,
contact the Thermo Fisher Scientific sales organization.
Sample rack/well plate detection sensor on shovel,
Barcode reader to detect rack type and shelf ID
Automatic stop when door is opened (interlock door switch)
Automatic opening of door after power failure
3 LEDs for status monitoring
All system parameters are logged in the Chromeleon Audit Trail.