Figure 3-7: Recorder Panel
The
Fixed Interval
is the recording interval for each group. The configuration of sensors in each group will influence on
the available recording interval. If you change the configuration of one sensor and this cause a longer processing time
Fixed Interval
the
might be changed.
Each Sensor will do a calculation based on the configuration for each recording interval.
If Acoustic Wave is enabled then the recording interval for the group containing the DCPS will be the wave
measurement period set by
current measurement period.
One recording interval will then consist of a wave measurement period followed by one current measurement period.
During the Wave measurement period only a reduced current measurement is calculated refer chapter
sensors in the same group will present one measurement during each recording interval.
Figure 3-8: Typical Recording Interval Acoustic Wave and Current
July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor
Wave Integration Time
System Configuration
in the
Recording Interval 30 min
Wave Measuring Time 20 min
Current Measurement Time 10 min
If the instrument is recording, under
Recorder
Panel, press
"Stop All
Groups".
Control Panel
The
is controlling the
SeaGuardII/SmartGuard recorder and not
individual sensors. However the Recorder
Panel consist of three recording groups
where you might put individual sensors in
each group to run them on different
recording interval. See Operation Manual
SeaGuardII / SmartGuard for more info.
Under each group you can start the group
immediately by
Start
Now, or start at a
later time using the
Start Delayed
setting, refer chapter
option.
3.6.2
and the
3.6.8.