Doppler Current Profiler Sensor
DCPS 5400 / 5402 / 5403
TD 304 OPERATING MANUAL
Doppler Current Profiler Sensor
July 2019

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DCPS 5400 and is the answer not in the manual?

Questions and answers

Summary of Contents for Aanderaa DCPS 5400

  • Page 1 TD 304 OPERATING MANUAL Doppler Current Profiler Sensor July 2019 Doppler Current Profiler Sensor DCPS 5400 / 5402 / 5403...
  • Page 2 25 May 2019 New Acoustic Wave Software, new Cell States description. New structure and reorganization of chapters, from image 20.4.26 Edition 20 July 2019 Updates and added chapter 9 to 11 from image 20.4.x © Copyright: Aanderaa Data Instruments AS...
  • Page 3 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 3 Contact information: Aanderaa Data Instrument AS Visiting address: TEL: +47 55 604800 PO BOX 103, Midttun Sandalsringen 5b FAX: +47 55 604801 5843 Bergen, NORWAY 5843 Bergen, Norway EMAIL: aanderaa.support@xyleminc.com...
  • Page 4: Table Of Contents

    Page 4 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Table of Contents Introduction ............................8 Purpose and scope..........................8 CHAPTER 1 Short description and specifications of the DCPS ............ 11 1.1 Description ............................11 1.2 Sensor Dimension ........................... 13 1.3 Sensor pin configuration and data output..................
  • Page 5 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 5 2.4.12 Output enabling – System Parameters ..................56 2.4.13 Output enabling – Virtual Sensors ................... 56 2.5 User Maintenance settings ......................57 2.5.1 Mandatory ..........................57 2.5.2 Site Info .............................
  • Page 6 Page 6 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 3.7.2 Site Info ............................. 91 3.7.3 Serial Port ..........................91 3.7.4 Licensed Options ........................92 3.7.5 Wave Measurement ........................92 3.7.6 Profiler Dependencies ....................... 93 3.8 Sensor software versions for use together with SeaGuardII / SmartGuard ........93 CHAPTER 4 Logging data via AADI Real-Time Collector .............
  • Page 7 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 7 8.1 General information ........................127 8.2 Parameter list ..........................127 8.3 Parameter calculation ........................127 8.4 Other wave descriptions ........................ 129 CHAPTER 9 Other details ......................130 9.1 Compass heading 0°...
  • Page 8: Introduction

    Profiler Sensor 5400/5402/5403-series. These sensors are described in a single manual since the measurement principle and electronics are the same for all sensors. The DCPS 5400 is the shallow water version with a maximum depth rating of 300 meters. The DCPS 5402 is the intermediate water version with a maximum depth rating of 4500 meters, and the DCPS 5403 is the deep water version with a maximum depth rating of 6000 meters.
  • Page 9 Pressure Certificate D-409 Data Sheet SeaGuardII DCP D-422 Data Sheet SeaGuardII DCP Wave D-411 Data Sheet DCPS 5400/5400R, 5402/5402R, 5403/5403R TD 310 Theoretical Primer for Doppler Current Profiler TD 303 Manual for SeaGuardII Platform TD 312 Configuration guide for DCPS and SeaGuardII DCP...
  • Page 10 Page 10 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Abbreviations AiCaP Aanderaa Protocol: Automated idle Line CANbus Protocol ASCII American Standard Code for Information Interchange Controller Area Network - sometimes referred to as CANbus COM port...
  • Page 11: Chapter 1 Short Description And Specifications Of The Dcps

    CHAPTER 1 Short description and specifications of the DCPS 1.1 Description The Aanderaa Doppler Current Profiler Sensor is an acoustic current profiling sensor based on the Doppler shift principle. It can operate either in narrowband or in broadband mode; the two measurement techniques are described in TD 310, Primer for Doppler Current Profiler.
  • Page 12 Record State for each measurement output. These status parameters indicate if the conditions have been good or not. The DCPS 5400 sensor (300 meter version) has a POM Housing and titanium base plate while DCPS 5402 and DCPS 5403 also have titanium housing. The DCPS 5403 has a higher titanium grade (grade 5) than DCPS 5402 (grade 2) and withstand higher pressure.
  • Page 13: Sensor Dimension

    Figure 1-2: Sensor Dimension IW/DW version 1.3 Sensor pin configuration and data output DCPS 5400/5402/5403 are using a 10-pin lemo plug for RS-232 and AiCaP communication between sensor and logger; for pin configuration. DCPS 5400R/5402R/5403R is using a 10-pin lemo plug for RS-422 connection to Figure 1-3 external logger;...
  • Page 14: Sensor Pin Configuration 5400R/5402R/5403R With Rs-422 Communication

    Page 14 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Sensor Signal name Input (I) Sensor plug Output (O) MCBH10M. Pin no: Positive supply CAN_H CAN_L BOOT ENABLE Table 1-1: 5400/5402/5403 Sensor pin configuration 10-pin plug AiCaP and RS-232 1.3.2 Sensor pin configuration 5400R/5402R/5403R with RS-422 communication Figure 1-4 5400R/5402R/5403R Sensor communication 10-pin plug with RS-422 Signal name...
  • Page 15: Sensor Connection

    Page 15 1.4 Sensor Connection Aanderaa offers a wide range of cables for different use of the sensors, both standard cables for use with loggers using AiCaP, RS-232 and RS-422 but also special customer specified cables for use in project. See for an chapter 11.2...
  • Page 16: User Accessible Sensor Properties

    Page 16 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 1.5 User accessible sensor properties All configuration settings that determine the behaviour of the sensor are called properties and are stored in a persistent memory block (flash). One property can contain several data elements of equal type (Boolean, character, integer etc.). The different properties also have different access levels.
  • Page 17: Sensor Properties

    8 for more details. To enable Acoustic Wave you need to order and enter the Acoustic Wave License Key. For older versions of DCPS a hardware upgrade might also be needed. Please contact aanderaa.sales@xyleminc.com assistance. All sensor properties are listed in chapters 1.7.1 through 1.7.4.
  • Page 18: Factory Configuration

    1.7.1 Factory Configuration All properties in this section are Read Only, not possible to overwrite for the user. Only certified Aanderaa service personal can alter these settings. The access level for reading the status of this properties are however different for each property, see table for more details.
  • Page 19 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 19 Set limits 1, 2, 3 and Discard Data for “Cell has weak signal, strength below weak limit for broadband”, default values are -40, - 44, -48 and -50. If value exceeds the limits a bit in Cell State will be Strength Limits set.
  • Page 20: Deployment Settings

    Page 20 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 1.7.2 Deployment Settings Deployment Settings contains settings for instruments metadata like position and owner, but also site dependent properties storing data that might influence the measurement. Value stored in these setting can be used as default values but might also be exchanged with real-time measured data if sensors are connected to the AiCaP bus.
  • Page 21 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 21 The temperature in degree centigrade. The value is used for calculation of Salinity, Sound speed and Density. Can be Temperature Float altered while sensor is running to compensate for variable temperature.
  • Page 22 Page 22 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor A list of all available Heading parameters with Deg.M as unit is Heading Parameter ENUM shown in the dropdown menu. Make sure that the correct heading parameter is selected. Only in AiCaP mode. Heading Alignment Offset added to the AiCaP external heading.
  • Page 23: System Configuration

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 23 1.7.3 System Configuration This group is used to control the sensor via properties for configuring communication with logger, sensor setup and parameter enabling and controlling the output from sensor. Some of the properties are only visible depending on the mode selected or if the function is enabled or not.
  • Page 24 Page 24 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Fixed Heading Float The fixed heading value is used if fixed heading is enabled. [Deg.M] No / Set to ‘No’ if used upward looking and ‘Yes’ if used downward looking. Enable Upside Uses different calibration coefficient set depending on setting.
  • Page 25 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 25 Enables surface reference for Column 1. If not selected the column will C1 Enable be instrument referenced. Note! The sensor needs a pressure input to BOOL Surface compensate for tidal variations.
  • Page 26 Page 26 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor The distance to centre of first cell for Column 3 in meter, distance from surface if surface referenced or sensor if not surface referenced. C3 Distance ENUM Always sensor referenced if used upside down.
  • Page 27 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 27 Different groups of parameters that will control the output from sensor. If User Specified is selected you may select the individual parameter under User Specified – Profile Parameter. For all other selections the Select Profile ENUM individual setting will be overruled by the group setting.
  • Page 28 Page 28 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Enables cross difference output for all cells. Difference between Doppler shift measured by the transducers on same axis, for each Cross depth the speed in beam 1 - speed beam 3 + speed beam 2 – speed ENUM Difference beam 4 should be close to 0.
  • Page 29 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 29 Enables output of the measured input voltage internally in the sensor. Voltage This is a way to see if the input supply voltage starts dropping. ENUM Output ‘Off’...
  • Page 30 Page 30 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor (32-bit where 32 flags indicates different statuses) with a ‘1’ if there is a mismatch between the upside down setting and the orientation measured by the sensor (binary bits xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx1, where x are the other bits indicating other statuses).
  • Page 31: User Maintenance

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 31 1.7.4 User Maintenance This group contains sensor settings that normally are not altered by the user. To access most of these properties you need to send passkey(1000) or with Real-Time Collector use password: 1000. These properties are used to configure serial port settings, communication to and from sensor.
  • Page 32: Manufacturing And Quality Control

    1.8 Manufacturing and Quality Control Aanderaa Data Instruments products have a record for proven reliability. With over 50 years’ experience producing instruments for use in demanding environments around the globe you can count on our reputation of delivering the most reliable products available.
  • Page 33: Chapter 2 Stand-Alone Sensor Configuration Using Aadi Real-Time Collector

    RS-232 or RS-422 via the PC COM-port. The menus shown here are slightly different from the menus shown when the sensor is working in AiCaP mode and configured through an Aanderaa Datalogger via serial or USB connection to the PC (described in CHAPTER 3). It’s also possible to use terminal software to configure a stand-alone sensor, refer CHAPTER 5, but then you need to enter all the commands and send them to the sensor.
  • Page 34 Page 34 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor . First enter a new Refer Figure 2-2 Connection Name, we recommend using name and serial number. Select Serial Port in the dropdown menu under Port Setting, and choose the correct COM-port on your computer under...
  • Page 35: Establish Connection

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 35 2.1.1 Establish connection The new connection is now shown in the AADI Real-Time Collector connection list. Click on the line with the new connection to highlight the line Open Port and then click on the button.
  • Page 36: Control Panel

    Page 36 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 2.1.2 Control Panel In the Control Panel window you will find two tabs, Recorder Panel and Device Configuration. Under Recorder Panel Stop Recorder you can start and stop recordings. If the recorder is running first click on the as you are not allowed to configure the sensor when recording.
  • Page 37: Changing Values

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 37 The device configuration is separated into Deployment settings, System Configuration, User Maintenance, System overview Save configuration to file. You can save current settings to a backup file by pressing Save…...
  • Page 38 Page 38 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor In the next picture a list of all properties that has been changed will show up. If the list of configuration changes is correct press Next to start the update process.
  • Page 39: Deployment Settings

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 39 When the updating process is finished a confirmation will Finish show up. Press to continue. Figure 2-12 Configuration Update Completed Deployment Settings Deployment Settings is separated in 5 groups.
  • Page 40: Common Settings

    Page 40 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 2.3.1 Common Settings Figure 2-14: Common Settings The interval can be changed in the Deployment Settings but can also be defined in the Control Panel in the Recorder Panel before starting the sensor.
  • Page 41: Fixed Settings

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 41 Enable Magnetic Declination is normally used when sensor is used close to South Pole or North Pole. Declination Angle (Deg.M) is a value to correct for the magnetic variation on the site where the sensor is used. This is the angle in degrees between magnetic north and true north.
  • Page 42: Calculations

    Page 42 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 2.3.5 Calculations Figure 2-18: Calculations Under Calculations you find only one Property, Enable Derived Sound Speed. If enabled the sound speed will be calculated from the other available sensors (only in AiCaP mode) or by using fixed values. This overrides the Sound Speed setting.
  • Page 43: System Configuration

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 43 System Configuration System Configuration separated in 13 groups. The settings in these groups are mainly used to configure the sensor and control the output/ storage of data. Refer chapter for more 1.7.3...
  • Page 44 Page 44 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor In the second half of the System Configuration menu you will find mainly setting that will control the output User Specified is selected under “Output – Profile Parameters>Select Profile Parameters then the individual...
  • Page 45: Common Settings

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 45 2.4.1 Common settings Figure 2-21: Common settings configuration Refer to : the communication protocol has to be defined under “Mode”. There are three different choices: Figure 2-21 AADI Real-Time is the correct mode (protocol) when used together with Real-Time Collector.
  • Page 46: Orientation Settings

    Page 46 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 2.4.3 Orientation settings Figure 2-23: Sensor orientation settings Enable Tilt Compensation is recommended to set active at all time because then the sensor will automatically compensate for any tilt. Enable Fixed Heading might be used if there is any magnetic disturbance that might interfere with a compass heading and the instrument has a fix position.
  • Page 47 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 47 The two methods have their own advantages: Broadband gives a lower single ping standard deviation which requires fewer ping (lower energy usage) to get Broadband a good measurement result. The disadvantage in is ambiguity.
  • Page 48: Surface Settings - Configuration

    Page 48 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 2.4.5 Surface settings – Configuration Figure 2-26: Surface cell configuration Enable Surface Cell will set a cell in the surface layer. The data is set as not valid if the sensor is placed upside down with the surface cell enabled.
  • Page 49: Column 2

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 49 Column 1 is always active (not possible to disable). Column 1 can have up to 75 cells. Column 2 can have up to 50 cells and Column 3 can have up to 25 cells (150 cells all together). Columns 2 and 3 have to be enabled to be used. C1 Enable Surface Reference: If set the cells will be referenced to the surface.
  • Page 50: Column 3

    Page 50 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor C2 Enable Column: If you want more columns. The columns can be different, mix of both reference and configuration. Typical combination is one column covering the whole water column, one surface referenced with high resolution near surface and one instrument reference near bottom.
  • Page 51: Output Enabling - Profile Parameters

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 51 2.4.9 Output enabling – Profile Parameters Figure 2-30: Output – Profile Parameters While configuring the sensor, it is possible to define which parameters should be calculated and sent from the sensor. Select Profile Parameters gives the user the choice to either select one of the predefined groups of Profile Parameter in User Specified –...
  • Page 52 Page 52 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Table 2-1: Profile Parameter additional parameters for each group Basic + Basic + Basic + Beam + Simple Basic Beam 3-Beam 3-Beam Full Output Output Output Output Output Output...
  • Page 53: User Specified Settings - Profile Parameters

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 53 Vertical Beam123 Vertical Beam124 3-Beam Combination Output (continue) Vertical Beam134 Vertical Beam234 SP Stdev AutoBeam Horizontal Speed AutoBeam AutoBeam Output Direction AutoBeam North AutoBeam East AutoBeam Vertical AutoBeam Point Parameters Noise Peak Level B1...
  • Page 54 Page 54 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor NE Speed Output outputs the calculated values for North Speed East Speed where speed in north and east direction gives a positive value and speed in South and West direction gives a negative value. 3-Beam Combination Output outputs the four 3-Beam solutions, Beam 123, Beam 134, Beam 124 and Beam 234 for SP Stdev Beam...
  • Page 55: Output Enabling - Sensor Parameters

    (Absolute tilt) is the tilt between the horizontal plane and the sensor’s plane. When upside down, the Abs tilt is varying around 180 degree. This is different from the Aanderaa RDCP (previous profiler generation). The new profiler is able to sense if it is upside down or not, and the absolute tilt indicates the actual direction of the sensor when it is running.
  • Page 56: Output Enabling - System Parameters

    Page 56 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 2.4.12 Output enabling – System Parameters The sensor can also output different system parameters like voltage, current draw while awake, and voltage to the acoustic transmitter circuits etc.
  • Page 57: User Maintenance Settings

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 57 User Maintenance settings User Maintenance, you find Under properties that are password protected and are set / altered by a trained user. It is not recommended to change properties unless instructed.
  • Page 58: Site Info

    Page 58 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 2.5.2 Site Info Figure 2-37: Site Info Owner is only used for information and will be a part of the Meta data from the sensor. Not used in any calculations. 2.5.3 Serial Port Interface...
  • Page 59: Terminal Protocol

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 59 2.5.4 Terminal Protocol Figure 2-38: Terminal Protocol Enable Old Time Setting. This setting enables old input in Smart Sensor Terminal mode for Interval and Polled pingrate. 2.5.5 Profiler Dependencies Figure 2-39: Profiler Dependencies...
  • Page 60: Chapter 3 Configuration Via Smartguard Or Seaguardii Datalogger

    CHAPTER 3 Configuration via SmartGuard or SeaGuardII Datalogger Introduction The Doppler Current Profiler Sensor 5400/5402/5403 can easily be installed on the Aanderaa SeaGuardII platform or connected using a cable. For more information about the SeaGuardII, refer to the TD 303, Operating manual SeaGuardII.
  • Page 61: Configuration With Real-Time Collector

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 61 Configuration with Real-Time collector 3.3.1 Connected via SeaGuardII • Connect the micro USB configuration cable to the USB connector in front of the instrument and to the PC •...
  • Page 62 Page 62 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor At first connection with AADI Real-Time Collector the connection list will be emty. Click on the button in the lower left corner to create a new connection NOTE: This only need to be done once.
  • Page 63: Control Panel

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 63 Refer to : AADI Real-Time Figure 3-5 Collector uses a default setting that fits for most Smart Sensor. However the DCPS sensor outputs a large amount of data and might have much longer response time (depending on the configuration) than other smart sensors.
  • Page 64 Page 64 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor If the instrument is recording, under Recorder Panel, press “Stop All Groups”. Control Panel is controlling the SeaGuardII/SmartGuard recorder and not individual sensors. However the Recorder Panel consist of three recording groups where you might put individual sensors in each group to run them on different...
  • Page 65 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 65 The device configuration is separated into Deployment settings, System Configuration, User Maintenance, System overview Save configuration to file. Settings related to the DCPS can be Deployment Settings, configured under System Configuration and User Maintenance System overview...
  • Page 66: Changing Values

    Page 66 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Changing Values To change the content of a property first tick in the value box of the actual property and then enter the text or number before pressing Next.
  • Page 67 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 67 An automatic process will start with 6 steps transferring and storing the new information/setting in the sensor Flash. If necessary a reset will be executed. Do not switch off before the entire process is completed.
  • Page 68: Deployment Settings

    Page 68 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Deployment settings Select Deployment Settings. DCPS Double-click on the to select the Doppler current profiler sensor (refer Figure 3-14) If additional sensors are connected to the SeaGuardII platform, it is possible to configure the DCPS to take input from these other sensors.
  • Page 69 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 69 Deployment Settings separated in 9 groups. The settings in this groups are mainly used as information or values used in internal calculations. Refer chapter 1.7.2 for more information on the properties in Deployment Settings If the surface cell, surface reference...
  • Page 70: Site Info

    Page 70 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 3.5.1 Site Info Figure 3-16: Site Info Site Info containing four properties: Location: Name of location where the instrument is deployed. • Geographic Position: GPS potion for deployment format Latitude, Longitude. •...
  • Page 71: Fixed Settings

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 71 3.5.3 Fixed Settings Figure 3-18: Fixed Settings Fixed Settings as shown in Figure 3-18 are settings used in internal calculation. Sound Speed (m/s) is the fixed sound speed setting in m/s. This value is used for cell positioning in the profile and for calculation of current speed values.
  • Page 72: Input Parameter Pressure

    Page 72 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 3.5.5 Input Parameter Pressure Figure 3-20: External pressure sensor parameter input Figure 3-20 gives an example of a wave & tide sensor connected to the Datalogger. The Wave and Tide sensor is selected as Pressure Sensor Id and the Tide Pressure is selected as...
  • Page 73: Input Parameter Conductivity

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 73 If the DCPS is running in a slow recording group in the multi-group recorder (more information about the Multi-group recorder is available in the TD 303 Chapter 3.3.1), it is possible to have the other sensors in a faster group to give more frequent update of input parameters.
  • Page 74: System Configuration

    Page 74 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor System Configuration Select System Configuration Device Configuration under (refer Figure 3-9). The different sensors connected to the same Datalogger will show up as selectable items, Refer Figure 3-25.
  • Page 75 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 75 System Configuration is separated in 14 groups. The settings in these groups are mainly used to configure the sensor and control the output/ storage of data. Refer chapter 1.7.3 for more information on the properties in System...
  • Page 76 Page 76 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor In the second half of the System Configuration menu you will find mainly setting that will control the output User Specified is selected under Output/Storage – Profile Parameters>Select Profile Parameters then the individual User Specified –...
  • Page 77: Orientation

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 77 3.6.1 Orientation Figure 3-28: Orientation Enable Tilt Compensation is recommended to set active at all time because then the sensor will automatically compensate for any tilt. Enable Fixed Heading might be used if there is any magnetic disturbance that might interfere with a compass heading and the instrument has a fix position.
  • Page 78: Current Profile Measurement

    Page 78 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Wave Integration Time is the integration time used to calculate waves. Number of samples divided by Wave Sampling Frequency Wave Integration Time. gives the minimum Wave Sampling Frequency is the ping frequency during wave measurement.
  • Page 79: Surface

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 79 Enable Burst Mode. The sensor can run in burst mode or spread mode. When burst mode is enabled the sensor performs all ping measurements at the beginning or end of the recording interval, dependent on the Burst Period Placement setting.
  • Page 80: Column 1

    Page 80 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 3.6.5 Column 1 Figure 3-33: Column 1 The sensor does not have any depth measurement itself. Be aware that you have to set correct fixed settings under Deployment settings if you want to use the Surface Cell or Surface Reference of columns.
  • Page 81: Column 2

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 81 C1 Number OF Cells: Available from 1 to 75 cells. The instrument can measure multiple cells along the column from instrument to the maximum range given by the scatter condition, or from surface and down to the instrument dependent on the configuration.
  • Page 82: Column 3

    Page 82 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 3.6.7 Column 3 Figure 3-35: Column 3 C3 Enable Column: Same as for C2 see chapter 3.6.6 C3 Enable Surface Reference: Same as for C1 see chapter 3.6.5 C3 Cell Size: Same as for C1 see chapter...
  • Page 83: Output/Storage Wave Parameters

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 83 3.6.9 Output/Storage Wave Parameters The Wave output parameters can be controlled by the user. Alternative options for each parameter are Off, Storage or Output+Storage where Off means the parameter is not calculated.
  • Page 84: Output/Storage Profile Parameters

    Page 84 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 3.6.10 Output/Storage Profile Parameters Figure 3-38: Output/Storage - Profile Parameters While configuring the sensor, it is possible to define which parameters should be calculated and sent from the sensor. Select Profile Parameters gives the user the choice to either select one of the predefined groups of Profile Parameter...
  • Page 85 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 85 Table 3-1: Profile Parameter additional parameters for each group Basic + Basic + Basic + Beam + Simple Basic Beam 3-Beam 3-Beam Full Output Output Output Output Output Output...
  • Page 86: User Specified - Profile Parameters

    Page 86 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Vertical Beam123 Vertical Beam124 3-Beam Combination Output (continue) Vertical Beam134 Vertical Beam234 SP Stdev AutoBeam Horizontal Speed AutoBeam AutoBeam Output Direction AutoBeam North AutoBeam East AutoBeam Vertical AutoBeam Point Parameters Noise Peak Level B1...
  • Page 87 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 87 NE Speed Output outputs the calculated values for North Speed East Speed where speed in north and east direction gives a positive value and speed in South and West direction gives a negative value. 3-Beam Combination Output outputs the four 3-Beam solutions, Beam 123, Beam 134, Beam 124 and Beam 234 for SP Stdev Beam...
  • Page 88: Output/Storage - Sensor Parameters

    (Absolute tilt) is the tilt between the horizontal plane and the sensor’s plane. When upside down, the Abs tilt is varying around 180 degree. This is different from the Aanderaa RDCP (previous profiler generation). The new profiler is able to sense if it is upside down or not, and the absolute tilt indicates the actual direction of the sensor when it is running.
  • Page 89: Output/Storage System Parameters

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 89 3.6.13 Output/Storage System Parameters The sensor can also output different system parameters like voltage, current draw while awake, and voltage to the acoustic transmitter circuits etc. that could be necessary during the QA&QC of data.
  • Page 90: User Maintenance

    Page 90 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor User Maintenance Under User Maintenance, you find properties that are password protected and are set / altered by a trained user. It is not recommended to change properties unless instructed.
  • Page 91: Mandatory

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 91 3.7.1 Mandatory Figure 3-44: Mandatory Node Description. All sensors are given a Node Description text like DCPS #xxx (xxx is the serial number of the sensor). The user can modify this node description text if required. Be aware that the node description changes to *Corrupt Configuration if it has lost the configuration in flash.
  • Page 92: Licensed Options

    Page 92 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 3.7.4 Licensed Options Figure 3-47: Licensed Options Acoustic Wave Product Number is only for factory use. Default value is The Acoustic Wave product number 5759. Acoustic Wave Option Key is used to enable the wave software.
  • Page 93: Profiler Dependencies

    Sensors with software versions lower than 8.2.4 should be upgraded to newer version. Newer software versions with stand-alone capabilities can also be used on sensors which are meant for use on SeaGuardII/ SmartGuard. This means that sensors with software versions from 8.1.25 can be used both on an Aanderaa Datalogger and as a stand-alone sensor.
  • Page 94: Chapter 4 Logging Data Via Aadi Real-Time Collector

    Page 94 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor CHAPTER 4 Logging data via AADI Real-Time Collector Logging data on PC The Real-Time Collector can save the incoming data to file, either to a txt-file or to xml-files. 4.1.1 Enabling file output If your connection is open (port...
  • Page 95 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 95 Then click on the “Advanced Settings…” button in the Connection Settings window; refer Figure 4-2 Figure 4-2: Connection settings menu File Output Choose from the list on the left side.
  • Page 96: Starting The Sensor And Logging To File

    Page 96 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 4.1.2 Starting the sensor and logging to file Click on the connection and “Open Port”. The Status turns green when the port is opened and connected. Figure 4-4: AADI Real-Time Collector start up menu Click on the “Control Panel…”...
  • Page 97: Viewing Incoming Data In Real-Time

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 97 Figure 4-6: Example of a txt-file obtained from the sensor using RT Collector Viewing incoming data in real-time When the sensor is running, the incoming data can be viewed under “Connection Logs…”...
  • Page 98 Page 98 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Figure 4-8: Visualization of incoming data from the sensor in real time Previous Entry Next Entry button. An Previous records or newer records can be viewed by clicking on button or automatic update to the last data message can be enabled by checking the Always show last entry check box.
  • Page 99: Chapter 5 Stand-Alone Sensor Configuration Using Smart Sensor Terminal

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 99 CHAPTER 5 Stand-alone Sensor configuration using Smart Sensor Terminal This chapter describes how to communicate with the Doppler Current Profiler sensor using the RS-232/RS-422 Smart Sensor Terminal protocol. Sensor versions and interface The 5400/5402/5403 sensors can be used on a SeaGuardII and SmartGuard (AiCaP) Dataloggers or connected to a RS-232 com-port (PC or other devices with RS232 com-port).
  • Page 100: Communication Setup

    Page 100 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Communication setup Most terminal programs can be used for Smart Sensor Terminal communication with the sensor when connected to a PC, e.g. Teraterm. The following Smart Sensor Terminal setup should be used: 115200 Baud 8 Data bits 1 Stop bit...
  • Page 101: Description Of Protocol

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 101 Any character will cause the electronics to return to normal operation; when the sensor has responded with the character ‘!’, new commands may be entered. When communicating with the sensor, you must start by pressing Enter. The sensor will respond in two ways (Comm TimeOut is 1 minute by default in the following description):...
  • Page 102: Save And Reset

    Page 102 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5.5.1 Save and Reset When the required properties are set, you should send a Save command to make sure that the new configuration is saved internally in the flash memory. The sensor always reads the configuration from the internal flash memory after Save reset and power up.
  • Page 103: Available Commands

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 103 5.5.2 Available commands Available commands and properties for the sensors are given in Table 5-1 Table 1-5 Table 1-8 respectively. Table 5-1: Available RS-232/RS-422 commands Command Description Start Start a measurement sequence according to current configuration...
  • Page 104: The Get Command

    Page 104 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5.5.3 The Get command command is used to read the value/values of a property and to read the latest value of a parameter. Property The command name followed by a returns a string in the following format: Property ProductNo SerialNo Value, ..., Value...
  • Page 105: Xml Commands

    XML-format. XML-output is a general format shared by all Aanderaa smart sensors; the output from different types of smart sensors can be read and presented as e.g. in a general smart sensor setup program.
  • Page 106 Page 106 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Column 1 is set up to 25 cells with 2m cell size. Since the Cell Center Spacing is 2m and equal to the Cell Size there will be no overlap and this gives a column which is 2m * 25 = 50m long.
  • Page 107: Help Command Output

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 107 Help command output Commands: Do_SUBCMD<CRLF> = Execute SUBCMD Get_PROPERTY<CRLF> = Output PROPERTY value. Get_All<CRLF> = Output all property values. Get_All Parameters<CRLF> = Output all parameter values. Set_PROPERTY(V,..V)<CRLF>...
  • Page 108 Page 108 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 4.0m,5.0m,6.0m,7.0m,8.0m,9.0m,10.0m,11.0m,12.0m,13.0m,14.0m,15.0m,16.0m,17.0m, 18.0m,19.0m,20.0m,25.0m,30.0m C3 Cell Size: 0.5m,1.0m,1.5m,2.0m,2.5m,3.0m,3.5m,4.0m,4.5m,5.0m C3 Distance First Cell Center: 1.5m,1.6m,1.7m,1.8m,1.9m,2.0m,2.1m,2.2m,2.3m,2.4m,2.5m,2.6m,2.7m,2.8m,2.9m, 3.0m,3.5m,4.0m,4.5m,5.0m,5.5m,6.0m,6.5m,7.0m,7.5m,8.0m,8.5m,9.0m,9.5m,10.0m,11.0m, 12.0m,13.0m,14.0m,15.0m,16.0m,17.0m,18.0m,19.0m,20.0m,22.0m,24.0m,26.0m,28.0m, 30.0m,32.0m,34.0m,36.0m,38.0m,40.0m,42.0m,44.0m,46.0m,48.0m,50.0m,55.0m,60.0m, 65.0m,70.0m C3 Number Of Cells: 1,2,3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25 C3 Cell Center Spacing: 0.1m,0.2m,0.3m,0.4m,0.5m,0.6m,0.7m,0.8m,0.9m,1.0m,1.1m,1.2m,1.3m,1.4m,1.5m, 1.6m, 1.7m,1.8m,1.9m,2.0m,2.1m,2.2m,2.3m,2.4m,2.5m,2.6m,2.7m,2.8m,2.9m,3.0m,3.2m,3.6m, 4.0m,5.0m, 0m,7.0m,8.0m,9.0m,10.0m,11.0m,12.0m,13.0m,14.0m,15.0m,16.0m,17.0m, 18.0m,19.0m,20.0m,25.0m,30.0m...
  • Page 109 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 109 The available sub commands Sample, Output, Do Refresh Do Calc MinTime) are listed next. All available values on the enumerated properties are also shown here. Some of the enumerations will change depending on configuration (Polled Pingrate property and Interval property).
  • Page 110 Page 110 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Example 3: Output from sensor when all selectable output is off //Press Enter to start communicating with the sensor, refer chapter ↵ //press Enter Stop ↵ // Stop current measurement.
  • Page 111 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 111 Figure 5-7: Example of the shortest output possible The data output message starts with MEASUREMENT followed by Product Number, Serial Number (tabulator Record between). The rest of the message is the output parameters which change from record to record. First there is a State which gives information about different quality parameters per record and then ping count per record.
  • Page 112: Output Parameters

    Page 112 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Output parameters 5.7.1 Parameters outside profile data These data are output first before the parameters from all cells in the columns. The name with unit shown in the table 3- 2 is the exact same way as these are sent from the sensor, i.e.
  • Page 113 SeaGuardII or SmartGuard the sensor can receive sensor input from other sensors to calculate Density, Depth, Salinity and Sound Speed in water (CTD input data). Without an Aanderaa SeaGuardII or SmartGuard the user has to send down the new settings to be able to get surface referred columns and surface cell. Usually the salinity does not change much during a short time period.
  • Page 114: Record State

    Page 114 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Figure 5-9: All parameters output outside profile data The order of these parameters is as follows: Heading, Std Dev Heading, Pitch, Roll, Abs Tilt, Max Tilt, Std Dev Tilt, Tilt Direction, Speed of Sound, Depth, Salinity, Density, Charge Voltage Vtx1, Charge Voltage Vtx2, Min Input Voltage, Input voltage, Input Current, Memory Used, Air Detect.
  • Page 115 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 115 Bit 8 xxxxxxxxxxxxxxxxxxxxxxx1xxxxxxxx Higher ambient noise measured on Beam 4 Bit 7,8 xxxxxxxxxxxxxxxxxxxxxxx11xxxxxxx Even higher ambient noise measured on Beam 4 Bit 9 xxxxxxxxxxxxxxxxxxxxxx1xxxxxxxxx High standard deviation heading Bit 10 xxxxxxxxxxxxxxxxxxxxx1xxxxxxxxxx Higher standard deviation heading Bit 9,10...
  • Page 116: Cell Parameters

    Page 116 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5.7.3 Cell Parameters All Cells have the same number of parameters; the parameters shown in Table 5-4 are repeated for each cell. The sensor can output from 1 cell in 1 column up to 150 cells total divided over three columns (75 in column 1, 50 in column 2 and 25 in column 3).
  • Page 117 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 117 Averaged signal from the four beams. The signal strength is calculated for each Strength[dB] Float beam for each ping and averaged at the end of the interval. Horizontal Speed Horizontal speed from the Autobeam solution.
  • Page 118 Page 118 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Calculated as (Beam1 Speed + Beam 3 Speed)-(Beam2 Speed+ Beam4 Speed) where the BeamX speeds are tilt compensated. The Cross Difference is Float Cross Difference[cm/s] calculated for each ping and averaged at the end of the interval. In a homogeneous water flow this value is close to zero.
  • Page 119 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 119 Set AutoBeam Output(Off/Output) and Set NE Speed Output(Off/Output) (both set to output SpStdev AutoBeam[cm/s] to be enabled). From software version 8.2.1: Set AutoBeam Output(Off/Output) together with Set Std Dev Speed Output(Off/Output) Strength[dB] Set Strength Output(Off/Output) Set AutoBeam Output(Off/Output) and Set AutoBeam Speed...
  • Page 120 Page 120 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Set NE Speed Output(Off/Output) and Set AutoBeam Output(Off/Output) From software version 8.2.1:Set AutoBeam Speed East AutoBeam[cm/s] Type(Polar/Rectangular/Polar+Rectangular/ Replace 4-Beam Data) and Set AutoBeam Output(Off/Output) set to Rectangular and Output to be enabled Set Vertical Speed Output(Off/Output) and Set 3-Beam Combination Output(Off/Output) (both set to output to be enabled).
  • Page 121: Cell States

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 121 5.7.4 Cell States Cell State 1 Cell State 2 is output for each cell giving QC data for each cell. Each bit has a status as shown in Table 5-6 Table 5-7.
  • Page 122 Page 122 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Binary: xxxx11xxxxxxxxxx 10 and Cell has High vertical current, above high vertical limit 3, default 30 cm/s Hexadecimal: xCxx Binary: xxx1xxxxxxxxxxxx Cell inside blanking zone Hexadecimal: 1xxx Binary: xx1xxxxxxxxxxxxx Cell inside illegible zone Hexadecimal: 2xxx...
  • Page 123 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 123 Binary: xxxxxxxxxx1xxxxx Beam2 cell out of range (too far away) Hexadecimal: xx2x Binary: xxxxxxxxx1xxxxxx Beam3 cell out of range (too far away) Hexadecimal: xx4x Binary: xxxxxxx1xxxxxxx Beam4 cell out of range (too far away) Hexadecimal: xx8x Binary: xxxxxxx1xxxxxxxx...
  • Page 124: Chapter 6 Software Versions And Stand-Alone Usage

    Page 124 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor CHAPTER 6 Software versions and Stand-alone usage The first software versions on the DCPS were made for use together with the SeaGuardII and SmartGuard Datalogger only. Using the sensor with a lower software version than 8.1.25 as a stand-alone sensor without a SeaGuard II or SmartGuard is not recommended and will not work properly.
  • Page 125: Chapter 7 Electro Magnetic Compatibility And Cables

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 125 CHAPTER 7 Electro Magnetic Compatibility and Cables In order for a manufacturer to legally produce and sell a product, it has to apply for CE marking. This means that the commercialized product is conform to the CE applicable standards and can freely circulate within the EFTA (European Free Trade Association) &...
  • Page 126: Emc Testing

    Page 126 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor EMC Testing The DCPS sensor has been tested at an accredited test laboratory to verify that the sensor fulfills the requirements in the EU EMC directive (89/336/EMC). Applied standards •...
  • Page 127: Chapter 8 Acoustic Wave

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 127 CHAPTER 8 Acoustic Wave General information Parameter list Table 8-1 Parameter list Output parameters Type Symbol Energy Spectrum Research E(f) Directional Spectrum Research Principal Directional Spectrum Research Orbital Ratio Spectrum Research...
  • Page 128 Page 128 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Principal Dir Spectrum, DWS is calculated based on the second order Fourier Coefficients. The principal wave direction has an ambiguity direction of 180 degree, but is forced to be in the same interval as the mean wave direction.
  • Page 129: Other Wave Descriptions

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 129 where: is the n order moment calculated from the Energy spectrum as; Other wave descriptions Wave Crest is the point on a wave with the maximum value or upward displacement within a cycle Wave Troughs is the point on a wave with the minimum or lowest point in a cycle...
  • Page 130: Chapter 9 Other Details

    Page 130 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor CHAPTER 9 Other details Compass heading 0° reference direction In some cases it is necessary to know the orientation of the sensor’s axes. This can be necessary when the sensor is placed in a fixed position with a fixed orientation (for example in a frame on the sea floor) where the magnetic influence by an object nearby gives an erroneous heading measurement.
  • Page 131: Orientation/Steering Pin As Reference

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 131 9.1.2 Orientation/Steering pin as reference Another method is to use the orientation pin underneath the sensor. This pin is placed between transducer 2 and 3. The heading is 225 degrees (360-135) if the orientation pin is towards magnetic north, i.e.
  • Page 132: Transducer 1 As Reference When Upside-Down

    Even alkaline batteries can be a problem if placed close enough to the sensor. When used on SeaGuardII the upper battery compartment can give problems when using alkaline battery cells. The battery cells provided by Aanderaa do not give problems when using the lower battery compartment. When batteries are own-built, you have to be aware of this problem.
  • Page 133: Checking The Tilt Sensor

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 133 Checking the tilt sensor The sensor has a 3-axis inclinometer. The tilt is converted to rotational angles pitch and roll. When the sensor is placed horizontal the pitch and roll should be close to zero. Pitch is the rotation around the y-axis while roll is the rotation angle around the x-axis.
  • Page 134: Chapter 10 Maintenance

    Page 134 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor CHAPTER 10 Maintenance With 50 years of instruments design and production for the scientific community, in use around the world, you can count on our reputation for designing the most reliable products available. We are guided by three underlying principles: quality, service, and commitment.
  • Page 135: Factory Service

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor Page 135 Factory service 10.3 Factory service is offered for maintenance, repair or calibration. When returning Doppler Current Profiler Sensor, always include the Instrument Service Order, Form No. 135; see our web pages under ‘Support and Training’.
  • Page 136: Example Of Test & Specifications Sheet And Certificates

    Page 136 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Example of Test & Specifications sheet and Certificates 10.4 Figure 10-1 Example of Test and Specification sheet...
  • Page 137 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Page 137 Figure 10-2: Example of Calibration Certificate compass and tilt page 1...
  • Page 138 Page 138 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Figure 10-3: Example of Calibration Certificate compass and tilt page 2...
  • Page 139 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Page 139 Figure 10-4 Example of Calibration Certificate for optional Temperature Sensor...
  • Page 140 Page 140 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Figure 10-5 Example of Pressure Certificate...
  • Page 141: Chapter 11 Installation

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400Page CHAPTER 11 Installation AADI Real-Time Collector connection with Windows 10 11.1 Windows 10 USB to SeaGuard / SmartGuard fixes Download Windows mobile device center (WMDC) 32 or 64 bit version: https://support.microsoft.com/en-us/help/931937/description-of-windows-mobile-device-center Windows 10 ver.
  • Page 142 Page 142 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Click ok and ok. Do the same with Windows Mobile-based device connectivity. Restart pc and plug in the USB cable from SmartGuard / SeaGuard.
  • Page 143: Connecting Cables

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Page 143 Connecting Cables 11.2 Aanderaa offers a wide range of standard cables; 11.2.1 For AiCaP and RS-232 use Table 11-1: Available cables for AiCaP and RS-232 •...
  • Page 144 8pin Subconn on 5340 cable • Needs 5340 cable for connection • 5414 Sensor Cable AiCaP • Free end cable for connection between sensor and Aanderaa logger • 4860 Sensor Cable RS-232 • Free end cable for connection between sensor and 3 party logger •...
  • Page 145: For Rs-422 Use

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Page 145 • 3855 Set-up and configuration cable • Non watertight Set-up cable with 9pin DsuB for connection between 10pin Lemo and PC for laboratory use only 11.2.2 For RS-422 use Table 11-2: Available cables for RS-422...
  • Page 146 Page 146 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 • 5159 Sensor Cable RS-422 • To be used with Real-Time cable • 5618 Set-up and configuration cable • Non watertight Set-up cable with 9pin DsuB for connection between 10pin Lemo and PC for laboratory use only •...
  • Page 147: Mounting Considerations

    July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Page 147 Mounting considerations 11.3 Figure 11-1: Inline mooring frame 5744A...
  • Page 148: Application Examples

    Page 148 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Application examples 11.4 DCPS is a flexible sensor that can be used in many different applications, either as a part of a SeaGuard II self- contained multiparameter observatory or as a part of a flexible user designed observation point.
  • Page 149 July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor 5400 Page 149 Example 2 in Figure 11-3 shows a more complex buoy system with meteorological sensors, satellite communication, solar panels as well as Motus directional wave sensor and DCPS. This system is also easily scalable with possibilities for adding water quality sensors and single-point current measurement for measuring current close to the surface.
  • Page 150 P.O.Box 103 Midtun, Sanddalsringen 5b, N-5843 Bergen, Norway Tel: +47 55 60 48 00 Fax: +47 55 60 48 01 email: aanderaa.info@xyleminc.com - www.aanderaa.com Aanderaa Data Instruments AS is a trademark of Xylem Inc. or one of its subsidiaries. © 2019 Xylem, Inc. July 2019...

This manual is also suitable for:

Dcps 5402Dcps 5403Dcps 5400rDcps 5402rDcps 5403r

Table of Contents