Aanderaa DCPS 5400 Operating Manual page 26

Doppler current profiler sensor
Hide thumbs Also See for DCPS 5400:
Table of Contents

Advertisement

Page 26
C3 Distance
ENUM
first Cell
C3 Number Of
ENUM
Cells
C3 Cell Center
ENUM
Spacing[m]
Enable
BOOL
Current from
Wave Ping
Wave Mean
Direction
ENUM
Output
Wave Mean
ENUM
Period Tm02
Output
Wave Energy
ENUM
Period Tm-10
Output
Energy
Spectrum
ENUM
Output
Directional
ENUM
Spectrum
Output
Principal Dir
ENUM
Spectrum
Output
Orbital Ratio
Spectrum
ENUM
Output
Fourier Coeff
ENUM
Spectrum
Output
Wave Depth
ENUM
Cell Output
July 2019 – TD 304 OPERATING MANUAL – Doppler Current Profiler Sensor
The distance to centre of first cell for Column 3 in meter, distance from
surface if surface referenced or sensor if not surface referenced.
1
Always sensor referenced if used upside down. Selectable from 1 to 70
meter. Default setting is 3 meter.
1
Number of Cells in Column 3, maximum 25 cells. Default setting is 10.
Spacing between two cells for Column 3 in meter. If spacing is equal to
cell size the two cells is next to each other. Spacing less than cell size
1
gives overlap and spacing more than cell size gives space between two
cells. Selectable from 0.1 to 30 meter. Default setting is 2 meter.
If enabled data from wave measurement will also be used to calculate
1
current. All current data are presented in Column 4. Only in AiCaP
mode and with Wave License Key.
Configuration of Wave Mean Dir Output
1
Off, Storage or Output+Storage. Only in AiCaP and with Wave License
Key.
Configuration of Wave Mean Period Tm02 Output
1
'Off, Storage or Output+Storage'. Only in AiCaP and with Wave License
Key.
Configuration of Wave Energy Period Tm-10 Output
1
'Off, Storage or Output+Storage'. Only in AiCaP and with Wave License
Key.
Configuration of Energy Spectrum Output
1
'Off, Storage or Output+Storage'. Only in AiCaP and with Wave License
Key.
Configuration of Directional Spectrum Output
1
'Off, Storage or Output+Storage'. Only in AiCaP and with Wave License
Key.
Configuration of Principal Dir Spectrum Output
1
'Off, Storage or Output+Storage'. Only in AiCaP and with Wave License
Key.
Configuration of Orbital Ratio Spectrum Output
1
'Off, Storage or Output+Storage'. Only in AiCaP and with Wave License
Key.
Configuration of Fourier Coeff Spectrum Output
1
'Off, Storage or Output+Storage'. Only in AiCaP and with Wave License
Key.
Configuration of Wave Depth Cell Output
1
'Off, Storage or Output+Storage'. Only in AiCaP and with Wave License
Key.
No /
SC
Low

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dcps 5402Dcps 5403Dcps 5400rDcps 5402rDcps 5403r

Table of Contents