6 testo 6610 probes
Accuracy (at 25 °C/77 °F)*
•
Humidity
◦
◦
◦
◦
•
Temperature
◦
* Refer to the charts below for the correlation between
temperature and accuracy.
Reproducibility
•
Better than ±0.2 % RH
Sensor
•
Response time without protective filter: t 90 max. 15 sec.
Probe dimensions
•
Diameter of probe shaft: 12 mm
•
L = approx. 200/500 mm
•
L – A = 165/465 mm
A see Overview of probe and filter types, page 74.
•
Probe length incl. probe shaft and filter
•
1/2/5/10 m
Pressure resistance
•
PN 10 (probe tip)**
** If installing probe under pressure, please use cutting ring screw
connection (order no. 0554 1795).
94
± (1.2 % RH + 0.007 x reading) for 0 to 90 % RH
± (1.6 % RH + 0.007 x reading) for 90 to 100 % RH
0.02 % RH/K, dependent upon the process temperature
(with a deviation of 25 °C/+77 °F)
0.02 % RH/K, dependent upon the electronics temperature
(with a deviation of 25 °C/+77 °F)
±0.15 °C (0.27 °F) with slope PT1000 class AA