7 Parameterizing, adjusting and analyzing software (P2A software)
Field
Limit values, relay 1 to 4/alarm
values, alarm 1 to 4
Relay x/alarm x
Not used
Collective alarm
Min control
110
Explanation
In this mask, the relays or display
alarms are parameterized.
Four relays or alarm values are
available (optional).
Relay is not used.
Hysteresis image and input
options are hidden.
A relay can be used as a collective
alarm detector in the event that
selected messages appear.
Selection of the messages (OR
linkage) by selecting the
checkbox.
If switched to ON (NO contact) or
OFF (NC contact) under the limit
value; in the event of a
subsequent exceeding of
value
plus Hysteresis, it is
switched to OFF (NO contact) or
ON (NC contact).