3-2-4
Setting the Data to Output after a Read
The code reader can be configured so that after a Read is executed, its read results are automatically
output to the TCP port it is connected to. Additional information such as print quality grade and code
position coordinates can be appended to the Read result output and the format of that output can be
modified.
Change the Read Result Output Condition
You can change the conditions by which you will output your Read results.
• WebLink - Setup - Gear Icon - Advanced Settings - I/O - Symbol Data Output
Setting Item
Symbol Data Output
Output Timing
Set the Data to be Output when There is a No Read.
You can change the data to output when there is a No Read result.
• WebLink - Setup - Gear Icon - Advanced Settings - I/O - No Read Message
Setting Item
No Read Message
Message
Autofocus Multicode Reader MicroHAWK V320-F/V330-F/V420-F/V430-F Series User Manual for Communication Settings (Z407-E1)
3 Controlling Operation and Data Output with Ethernet
Setting Value
Disabled
Disabled:
Match
Read result is not output.
Match:
Mismatch
The Read result is only output when it matches the Master
Good Read
Symbol set in the Matchcode function.
(Default)
Mismatch:
Only If All Are
The Read result is only output when it does not match the Mas-
Good Reads
ter Symbol set in the Matchcode function.
On Good Read:
Read results are output for even just one Good Read.
Only If All Are Good Reads:
The Read result is output only when all the symbols specified in
the Read Multiple Symbols function are successfully read.
As Soon As Pos-
As Soon As Possible:
sible
Outputs the Read result immediately on Good Read and ends
(Default)
the Read Cycle.
End of Read Cy-
End of Read Cycle:
cle
The Read result is not output until the End of Read Cycle con-
dition is met. The End of Read Cycle condition is set in
Advanced Settings - Read Cycle - End of Read Cycle.
Setting Value
Enabled
Enabled:
(Default)
A message is output when there is a No Read. However, if the
Trigger Mode set in Read Cycle is Continuous Read, no mes-
Disabled
sage is output regardless of this setting.
Disabled:
No message is output for a No Read.
NOREAD
Set the message to output when there is a No Read. You can set
(Default)
up to 64 ASCII characters.
Description
Description
3
3-33