Electrical Specifications - Vatech Green 21 PCT-90LH User Manual

Table of Contents

Advertisement

9.2

Electrical Specifications

Power supply voltage
Frequency
Power rating
The input line voltage depends on the local electrical distribution system.
Allowable input voltage fluctuation requirement: ± 10 %
Mode of operation: non-continuous operation (NFPA 70: long time operation) -
needs waiting time (at least 60 times the exposure time) before the next exposure
begins
Exposure time: Max. 18 s / Resting time: 10 min.
Column operation time: 1 min. / Resting time: 9 min.
Green21 (Model: PCT-90LH) Installation Manual
Item
AC 100 ~ 240 V
50 / 60 Hz
1.3 - 1.6 kVA
To assure line voltage quality, a separate 3-core
grounded power cable connected directly to central
distribution panel with an over-current circuit breaker
rated for 20/15A must be used.
The mains resistance should not exceed 100 V at 0.045-
ohm, 240 V at 0.19 ohm.
9. Technical Specifications
Description
101