9.2
Electrical Specifications
Power supply voltage
Frequency
Power rating
The input line voltage depends on the local electrical distribution system.
Allowable input voltage fluctuation requirement: ± 10 %
Mode of operation: non-continuous operation (NFPA 70: long time operation) -
needs waiting time (at least 60 times the exposure time) before the next exposure
begins
Exposure time: Max. 18 s / Resting time: 10 min.
Column operation time: 1 min. / Resting time: 9 min.
Green21 (Model: PCT-90LH) Installation Manual
Item
AC 100 ~ 240 V
50 / 60 Hz
1.3 - 1.6 kVA
To assure line voltage quality, a separate 3-core
grounded power cable connected directly to central
distribution panel with an over-current circuit breaker
rated for 20/15A must be used.
The mains resistance should not exceed 100 V at 0.045-
ohm, 240 V at 0.19 ohm.
9. Technical Specifications
Description
101