9. Device Specifications
9.3
Electrical Specifications
Tube Voltage
Tube Current
Exposure Time
Rated Voltage
9.4
During operating
Transport and storage
74
Item
The system will be available with a fixed tube voltage
specification based on the user selection.
Maximum exposures on a full charge: about 600 times at 65
kV, 2.5 mA, 0.2 s
Environmental Specifications
Item
Temperature
Relative humidity
Atmospheric pressure
Temperature
Relative humidity
Atmospheric pressure
Failure to follow the specifications above can result in severe
injury and device damage.
Description
Option 1: 60 kV
Option 2: 65 kV
2.5 mA (± 10 %)
0.05-1.0 s (± 3 % or 10 ms)
21.6 V
10 ~ 35 ℃
30 ~ 75 %
860 ~ 1060 hPa
-10 ~ 60 ℃
10 ~ 75 % non-condensing
860 ~ 1060 hPa
EzRay Air Portable (Model: VEX-P300) User Manual
fixed (± 5 %)
fixed (± 5 %)
dc
Description