Vatech Green 21 PCT-90LH User Manual page 63

Table of Contents

Advertisement

5.
CT - 4% cut align work
1)
Perform 4% cut according to the following criteria.
2)
Run C:\VCaptureSW\Util\CollimatorAlign\CollimatorAlign.exe file.
3)
Perform 4% cut align of FOV 21x19 of dental CT (see above criteria.)
4)
Set the Dental CT FOV21x19 collimator to its own criteria and click "CALC"
to set the 17x15, 12x9, 8x8 collimator.
5)
When the dental CT is completed, the ENT CT value is copied, calculated,
and set.
Green21 (Model: PCT-90LH) Installation Manual
CT: Threshold – 95
Collimator
Left (T-80%)
6 pixels
Right (T-80%)
4 pixels
Top (T-80%)
4 pixels
Bottom (T-80%)
4 pixels
Collimator value should be saved after proceeding to Capture
after clicking the Initial button, so Should shoot more than
once for each FOV.
Collimator value should be saved after proceeding to Capture
after clicking the Initial button, so Should shoot more than
once for each FOV.
When Spiral Artifacts occurs, conduct the step iii) and lengthen
the collimator interval.
* For designated countries, currently Korea and America, type
the Left and Right collimator position values as 1.0 less than
the current value.
4. Installation Procedures
Min
Max
8 pixels
6 pixels
10 pixels
10 pixels
51