Disabling the packing mode
1.
Click on the General tab in the control panel.
In the Networking option, select the Serial checkbox and set Serial port and
2.
Baud Rate as follows and then click on Manager.
3.
Enter the command PVER] to verify the current mode. Note that the equipment is
now in the packing mode.
Green21 (Model: PCT-90LH) Installation Manual
Green21 has a packing mode built into the system to prevent
the unit from being damaged while shipping and transporting.
Thus, it is in the packing mode by factory default. Disabling
the packing mode is required at this step for successful
installation.
Unless the packing mode is disabled, no operation will
happen even after the equipment is turned on.
If an error has occurred during connection, make sure the
COM port setting is correct as follows:
1) Run Device Manager.
2) Check the COM port assigned to CONTACT-US Virtual
Serial Port under Ports (COM & LPT).
3) If necessary, change the COM port in the Networking
option and click on Save.
Figure 7-35. Checking the packing mode