Important Precautions
1) Rotating Unit
vi
Never touch or hold the following areas while moving,
installing, or operating the equipment.
2) Handle Frame, Touch Screen, and
Column Unit
Installation Site
1. The PC monitor, emergency cut off switch and X-Ray
exposure switch should be installed near the operator so
that he or she can manage them simultaneously in an
emergency.
2. Proper shielding of the room is essential: Since these
requirements vary depending on the country, it is the
installer's responsibilities to verify that all applicable
radiation safety requirements are met.
3. This equipment should not be installed in the immediate
vicinity of other devices.
4. Do not install the equipment in an area that is exposed to
strong electromagnetic fields.
5. Do not install this system in an area where there is the risk
of an explosion.
6. The electrical installation of this system shall comply with all
local code requirements for electro-medical systems: IEC
60364-7-710:2002.
7. It is strongly recommended that a UPS be installed at the
same time as the equipment.
8. The equipment, PC, and all peripheral devices must be well
prepared.
Make sure to hold transfer handles during transport and
installation referring to Chapter 0. Installation Procedures.
Green21 (Model: PCT-90LH) Installation Manual