Environmental Conditions
4
Refer to
Safety information
This instrument is intended for "indoor use" only.
The following table shows the environmental requirements and the corresponding
characteristics for the product.
Environmental Requirements
Temperature
Maximum Relative Humidity
(non-condensing)
Altitude
Pollution degree*
* See table below for Pollution Degree definitions
Pollution Degree
Description
1
No pollution or only dry, non-conductive pollution occurs. The pollution has no influence.
Example: A clean room or climate-controlled office environment.
2
Normally only dry non-conductive pollution occurs. Occasionally a temporary conductivity
caused by condensation may occur. Example: General indoor environment.
3
Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes
conductive due to condensation which is expected. Example: Sheltered outdoor environment.
on page 77 for module-specific safety considerations.
General characteristics
Operating condition: 0°C to 40°C
Storage condition: -40°C to 70°C
Type tested, 95% RH up to 40°C, decreases linearly to 40% RH at 45°C
Operating condition: Up to 10,000 ft (3048m)
Storage condition: Up to 15,000 ft (4572m)
Pollution Degree 2
M5200A PXIe Digitizer Modules Startup Guide