Performance Tests
Check DC Voltage
Measurement Accuracy
1–38
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
WARNING. The generator is capable of outputting dangerous voltages. Be sure to
set the DC calibration generator to off or 0 volts before connecting, disconnect-
ing, and/or moving the test hookup during the performance of this procedure.
Also, check that the calibrator does not have shorting straps installed between
the DC and sense outputs or grounds.
Equipment
Two dual-banana connectors (Item 6)
required
One BNC T connector (Item 7)
One DC calibration generator (Item 9)
Two precision 50 W coaxial cables (Item 5)
Prerequisites
The oscilloscope must meet the prerequisites listed on page 1–17
See Input Channels versus Model on page 1–2.
1. Install the test hookup and preset the instrument controls:
a. Hook up the test-signal source:
H Set the output of a DC calibration generator to off or 0 volts.
H Connect the output of a DC calibration generator through a
dual-banana connector followed by a 50 W precision coaxial cable to
one side of a BNC T connector. See Figure 1–5.
H Connect the Sense output of the generator through a second
dual-banana connector followed by a 50 W precision coaxial cable to
the other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 1–5.
Output Sense
DC Calibrator
Dual banana to
BNC adapters
Figure 1–5: Initial test hookup
50 W coaxial cables
Oscilloscope
BNC T
connector