Performance Tests
3
Read results.
Set the generator
(reference) frequency to
1
the test frequency from
Table 1–4.
Set the horizontal scale
2
from Table 1–4.
1–46
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
TRIGGER LEVEL knob to do this. (Full bandwidth varies with
TDS model as is shown in Table 1–4).
g. Measure the test signal:
H Set the frequency of the generator, as shown on screen, to the test
frequency in Table 1–4 that corresponds to the vertical scale set in
substep d. See Figure 1–9.
H Set the horizontal SCALE to the horizontal scale setting in
Table 1–4 that corresponds to the vertical scale set in substep d.
Press SET LEVEL TO 50% as necessary to trigger the signal.
H Read the results at the CHx Pk-Pk readout, which will automatically
measure the amplitude of the test signal. See Figure 1–9.
Figure 1–9: Measurement of analog bandwidth
h. Check against limits:
H CHECK that the Pk-Pk readout on screen is within the limits listed
in Table 1–4 for the current vertical scale setting.
H Enter voltage on test record.
H When finished checking, set the horizontal SCALE back to the
50 ns (TDS694C 40 ns) setting.