Multiplexer
Start multiplexer measurement
Enter location ID
Diagnostic measurement at VIBRONET multiplexer
For machinery fault diagnosis you can conduct the following
measurement tasks at a multiplexer measurement location:
•
Vibration-based measurement types in Multimode
•
Only with 'Current Linedrive' acceleration sensor
•
1- or 2-channel
2-channel measurement tasks, such as Orbit, are possible only
with two string lines, since only one line is connected and can be
processed for each measuring channel.
For RPM-dependent measurements, such as order spectrum or
phase, use a suitable sensor (e.g. VIB 6.631), which you connect to
the VIBXPERT RPM channel.
To connect VIBXPERT to a multiplexer you need the following:
•
Connecting cable, VIB 5.346
•
BNC connecting adapter, VIB 5.346 MUX
Diagnostic measurement
•
Connect the VIBXPERT to the multiplexer (see p. 3-42).
•
Connect an RPM sensor to the VIBXPERT, if necessary.
•
In the startup screen, click 'Multimode' and mark the
corresponding measurement task icon.
•
Press the MENU button and click 'Execute on MUX'.
•
Configure the measurement location address:
- MUX ID = Number of the multiplexer in the line section
- MUX Input = Number of the measuring channel on the
multiplexer
•
To start the measurement, click 'Measure'.