Measurement tasks
Multitask (Multimode)
Selecting individual meas. tasks (left)
Setting the trigger parameters (right)
Measurement task
Evaluation setup
Multi-measurement task in the 'Multimode' operating mode
Multi-measurement tasks are created in the Task Manager:
• Open the Task Manager (see P. 3-19).
• Select the upper menu 'Measurement task' and press the MENU
key.
• Click on 'New' and enter a name for the new measurement task.
• In the 'Start measurement' field, define whether the
measurements should be started consecutively or simultaneously.
With the 'Synchronous' setting, you can additionally define a
trigger channel ('Synchronous + Channel ...').
The measurement task, sensors and trigger parameters can be set
up in the respective tabs:
Task tab
Here you can set the sensor for each channel (A/B), and define
whether the rpm should be recorded as well.
Channel A / Channel B tab
Here you can define the measurement task for each channel:
• Press the MENU key and click on 'Select task' (see below). The
import wizard appears, in which you can select the individual
measurement tasks in three steps (see also the next section, 'Dual
measurement').
Trigger tab
The trigger settings only apply when the multi-measurement task
is to start with a trigger (see 'Start measurement' above). The first
measurement starts as soon as the Trigger level is exceeded and
the Trigger start time has elapsed (see also p. 3-22). The trigger
levels relates to the measurement range.