Entering the normalization parameters
Before the shock pulse measurement starts, the normalization
screen appears:
• Select the type of normalization:
No Normalization: Shock pulse value specified in dBsv.
Manual: RPM is entered manually before the shock pulse
measurement.
External: RPM is measured with an RPM sensor before the shock
pulse measurement.
• Enter the dBa value if it is known. Otherwise, carry out a
normalization adaptation (set dBa to '0', see previous section).
• Enter the diameter of the bearing inner race.
• If the normalization type is set to 'manual', enter the RPM (see
below).
• To start the shock pulse measurement, click on 'OK'.
Measurement tasks
Shock pulse - Normalization
5-3