Changing The Measurement Task (M); Setting Up A New Measurement Task (M) - Fluke VIBXPERT II User Manual

Data collector & vibration analyzer
Hide thumbs Also See for VIBXPERT II:
Table of Contents

Advertisement

Use this option when you do not have an RPM sensor on the route
and need to perform measurements that require the RPM (e.g. roller
bearing condition, order-based spectra). If the measurement task
already has information on the RPM, this option is ignored.

Changing the measurement task (M)

• Highlight the measurement type in the selection screen.
• Press the F key. A list appears with all the measurement tasks set
up for this measurement type.
• Click on the required measurement task (see below).

Setting up a new measurement task (M)

VIBXPERT has a large selection of frequently used measurement
tasks* that can be extended by user-defined measurement tasks
if necessary. To configure a new measurement task, proceed as
follows:
• Highlight the measurement type in the selection screen.
• Press the MENU key and click on 'Task Manager'. The
Task Manager appears and can be used to set up the new
measurement task (see P. 3-2).
• Click on the 'Measurement task' pulldown menu.
• Press the MENU key and click on 'New'. The text editor appears.
• Enter a name for the new measurement task.
• Change the measurement channel if necessary.
• Select the setups for measurement (see below), sensor**,
evaluation or RPM in the 'Setup' field of the pulldown menu.
If the matching entry cannot be found, a new setup must be
created (see next section).
• Then press the MENU key and click on 'OK'.
Options
M = in Multimode only
* Preconfigured measurement tasks/
setups are marked with a lock and
cannot be edited (except for
transducer & channel).
Open the 'Task Manager '
**Initially only the 'available transduc-
ers' appear in the transducer setup (see
P. 2-19). To display all transducers,
click on the bottom list entry '...more
transducers'
Left:
Changing the measurement task
Right:
Task Manager
Setups for
measurement, sensor, evaluation and
RPM measurement
3-19

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vib 5.310

Table of Contents