NA (Network Analyzer) Mode
NA Mode Settings
— Press Enter

Single or Continuous Measure

This setting determines whether the FieldFox sweeps continuously or only once
each time the
or to allow you to save or analyze a specific measurement trace.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
— Press Sweep 3.
— Then choose one of the following:
— You can also use Run / Hold +/- to toggle between Single and Continuous.

Resolution (Number of Data Points)

Data points are individual measurements that are made and plotted across the
X-axis to form a trace. Select more data points to increase measurement
resolution. However, more data points also takes more time to complete an
entire measurement sweep.
When the Resolution is changed after a calibration is performed, the cal
becomes interpolated. Learn more in
How to set Resolution
— Press Sweep 3.
— Then Resolution.
— Then choose from the following:
— Using SCPI, Resolution can be set to ANY number of points between 3 and
100
Single
button is pressed. Use Single to conserve battery power
Single
Automatically sets Continuous OFF and causes FieldFox to
make ONE sweep, then hold for the next Single key press. Hold is
annotated in the upper left corner of the display when NOT sweeping,
and changes to an arrow --> while the sweep occurs.
Continuous
Makes continuous sweeps. This is the typical setting
when battery power is not critical.
101 |201
|
401
|
601
|
801
10001. See the Programming Guide at
http://na.support.keysight.com/fieldfox/help/Programming/webhelp/FFP
rogrammingHelp.htm
"Interpolation *" on page
|
1001
|
1601
|
4001
| 10001.
Keysight N9927-90001 User's Guide
140.