Technical Service; Accessories/Spare Parts - Braun Aesculap Acculan 4 Instructions For Use/Technical Description

Dermatome ga340 / dermatome 0.1 mm ga341
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
en
Issue
Dermatome blade not
moving
Insufficient cutting per-
formance of dermatome
blade
Cutting depth
uneven/deviating from
setting

10. Technical Service

Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
WARNING
Do not perform any servicing or maintenance
work under any circumstances while the product
is being used on a patient.
Do not modify the product.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
For
service
and
repairs,
B. Braun/Aesculapagency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
14
Cause
Defective gearings
Dermatome blade blunted
Adjusted dermatome
Skin grafts uneven / too thick /
too thin
please
contact
your
Detection
Product runs loudly
Worn cutting edge of dermatome blade
Product has insufficient power
Mechanical influence (e.g. fall on floor)

11. Accessories/Spare parts

Art. no.
GA059
GA346
GA340345
GA643817
GA670210
GA675
national
GA678
GA679
GB228R
GB256R
GB487R
GB488R
GB489R
GB498R
GB600
TA014548
TA014549
Remedy
Have the manufacturer repair the product
Replace dermatome blade
Have the manufacturer repair the product
Have the manufacturer repair the product
Designation
STERILIT-Power-Systems drip lubricator (50 ml)
Li-Ion battery short
Nut
Flap rod
Flap
Lid
Sterile funnel
Battery removal device
Dermatome blade
ECCOS set for dermatome
ECCOS fixation for lid
ECCOS fixation for sterile funnel
ECCOS fixation for battery removal device
ECCOS fixation for straight machine
STERILIT Power Systems oil spray
Instructions for use of Dermatome GA340/GA341
(A4 for ring binder)
Instructions for use of Dermatome GA340/GA341
(flyer)

Advertisement

Table of Contents