7.13 Storage

Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
8.

Maintenance

To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance labeling or at least once a year.
YYYY-MM
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
9.

Troubleshooting list

Note
For additional information, see the operating instructions for the
ELAN 4 electro control unit GA800 (TA014401).

10. Technical Service

DANGER
Risk of death to patients and users in case of malfunction and/or fail-
ure of protective measures!
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
Do not modify the product.
CAUTION
Modifications of medical devices may result in a loss of potential
guarantee/warranty claims and forfeiture of applicable licenses.
Do not modify product.
Contact national B. Braun/Aesculaprepresentative for service and
repair.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
10
Document No.: TA014436 - Version: 1.0 - Document ID: SOP-AIC-5002244 Date/Time Printed/Viewed: 2022-04-13 16:01 (CET)

11. Accessories/Spare parts

Art. no.
Designation
GA344244
Kirschner wire protection sleeve
GA344211
Rinsing adapter
GB072R
ECCOS winding aid for cables and hoses
GB073R
ELAN 4 electro ECCOS motor cable
GB262R
ECCOS tray with fixations for GA844
GB600
STERILIT Power Systems oil spray
GB600880
Oil spray adapter for GA344/GA844
TA011944
Cleaning brush
GB482R
ECCOS fixation for GA344/GA844
TA014401
Instructions for use for ELAN 4 electro control unit
GA800 (A4 for ring-binder)
TA014436
Instructions for use for ELAN 4 electro drill tool
GA844 (A4 for ring-binder)
TA014437
Instructions for use of ELAN 4 electro drill GA844
(flyer)
TA014452
Instructions for use for attachments for the small
drill tool GA344 and the drill tool GA844 (A4 for
ring-binder)
TA014453
Instructions for use for attachments for the small
drill tool GA344 and the drill tool GA844 (flyer)

12. Technical data

12.1 Classification acc. to Directive 93/42/EEC

Art. no.
Designation
GA844
ELAN 4 electro drill

12.2 Performance data, information about standards

Max. power
approx. 200 W
Speed
0 rpm to max. 1,250 rpm ± 5 %
Rotational direction
Right and left rotation, oscillation
Cannulation
3.3 mm
Weight
0.8 kg ± 10 %
Cable length
4 m ± 10 %
Dimensions (L × W × H)
122 × 130 × 27 mm ± 5 %
Applied part
Type BF
EMC
IEC/DIN EN 60601-1-2
Conforming to standard
IEC/DIN EN 60601-1
After 350 preparatory cycles, the product was tested by the manufacturer
and passed.
Effective
Class
IIa