Aesculap
Oscillating saw GA331
Issue
Cause
Insufficient cutting
Saw blade teeth worn
performance of the
Battery too weak
saw blade
Battery worn/defective
Saw blade overheated
The saw blade cannot
Coupling components or saw
be coupled
blades deformed
-
Saw blade cannot be
Locking pin not fully releasing
uncoupled
the saw blade

10. Technical Service

Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
WARNING
Do not perform any servicing or maintenance
work under any circumstances while the product
is being used on a patient.
Do not modify the product.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
For
service
and
repairs,
B. Braun/Aesculapagency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
14
Acculan 4
®
please
contact
your
Detection
Blunt saw blade
Insufficient motor power/speed
Inadequate chip removal
Deformation
Incompatible saw blade
Button for saw blade release not
pressed completely

11. Accessories/Spare parts

Art. no.
GA675
GA676
GA678
GA679
GB243R
GB244R
national
GB600
GB495R
GB487R
GB488R
GB489R
TA014542
TA014543
Remedy
Replace saw blade
Charge battery
Replace battery
Move the saw back and forth during sawing and
ensure sufficient irrigation of the saw blade
Have repaired by manufacturer
Insert a compatible Aesculap saw blade, see
Accessories/Spare parts
Fully press the saw blade release button, see
Attaching and detaching the saw blade
Designation
Lid
NiMH battery long
Sterile funnel
Battery removal device
ECCOS sieve basket with fixation for Acculan 4
STERILIT Power Systems oil spray
ECCOS fixation for machine
ECCOS fixation for lid
ECCOS fixation for sterile funnel
ECCOS fixation for battery removal device
Instructions for use for oscillating saw GA331 (A4
for ring binder)
Instructions for use for oscillating saw GA331
(flyer)

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ga331

Table of Contents