Using A Dap Meter - AGFA DR 100e User Manual

Hide thumbs Also See for DR 100e:
Table of Contents

Advertisement

Using a DAP meter

The function of the DAP meter is to measure the dose-area product (μGy.m
that is going out towards the patient, accumulated over all exposures.
1. Before starting the examination of a patient, reset the DAP value by
pressing the F1 and reset button simultaneously.
Figure 29: Reset the DAP value
2. Perform the examination by performing all exposures for the patient.
After each exposure, the DAP value is incremented by the measured dose.
3. Read the accumulated DAP value on the first line of the display.
Figure 30: DAP value
CAUTION:
The installation and maintenance of the DAP meter can be
performed only by authorized service personnel.
DR 100e (Film/CR) | Advanced Operation | 65
2
)
0360B EN 20181015 0853

Advertisement

Table of Contents
loading

This manual is also suitable for:

6012/0016012/002

Table of Contents