28 | DR 100e (Film/CR) | Introduction
Message
INVERTER KV ERROR
INV. OVERLOAD
INVERTER ERROR
TUBE CALIB. ERR.
XR HANDSWITCH ERR. Faulty exposure button. Check the integrity of
0360B EN 20181015 0853
Description
During the exposure,
the kV value has de-
creased under 75% or
increased over 110% of
the requested value.
Unbalance in the high
voltage circuit during
the exposure
Power to the inverter is
out of range.
IGBT drivers error.
Error of X-ray tube cali-
bration.
Action
Press the reset button
and repeat the expo-
sure.
Press the reset button to
continue.
Press the reset button to
continue.
Call Service.
the exposure button,
switch off the device
and switch it on, try
again, if the error per-
sists call Service.