Anatomic program exposure parameters

The default X-ray exposure parameters for each exposure type can be
customized. The device is delivered with a set of X-ray exposure parameters
that are intended as a guideline. The default X-ray exposure parameters are
only starting points that can be replaced by more specific X-ray exposure
parameters developed by the operator.
In the following table the dose values are expressed in µGym
Product), the exposures are set with standard values for clinical investigations
on the patient and performed with the X-ray tube with a SID of 100 cm and
without grid.
The dose measurement has been performed according to IEC 60601-1-3 §
5.2.4.2 with the dosimetric chamber supplied with the equipment and the
collimator completely open.
Table 6: Anatomic group for examinations of the body
Anatomic
program
Chest AP
Abdomen AP
Abdomen
LAT
Chest LAT
Sternum
Ribs
Table 7: Anatomic group for examinations of the skull/spine
Anatomic
program
Skull AP
Thora. Spine
AP
Lumbar
Spine AP
Cervical
Spine
Focus
kV
Large
110
Large
81
Large
90
Large
110
Small
85
Small
70
Focus
kV
Small
77
Small
77
Large
81
Small
66
DR 100e (Film/CR) | Product Information | 83
mAs
2
16
20
4
4
6.3
mAs
10
20
20
8
2
(Dose Area
2
µGym
17.83
76.37
118.94
34.63
22.84
24.12
2
µGym
46.71
93.97
104.10
27.07
0360B EN 20181015 0853