94 | DR 100e (Film/CR) | Product Information
Harmonic emis-
sion in accord-
ance with IEC
61000-3-2
Voltage fluctua-
tions / flickering
in accordance
with IEC
61000-3-3
The device is used in a professional healthcare / radiological environment.
Environmental conditions are stated in the user manual.
This device was tested for a professional healthcare environment as described
above. Nevertheless the HF-emission and immunity can be influenced by
connected data cables depending on length and the manner of installation.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Resistance to Jam-
ming Test
Discharge of static
electricity in accord-
ance with IEC
610004-2
Fast transient electri-
cal disturbance varia-
bles / bursts in accord-
ance with IEC
610004-4
Impulse voltages
(surges) in accordance
with IEC 61000-4-5
Voltage breakdown,
short term interrup-
tions and variations in
the voltage supplied in
0360B EN 20181015 0853
Class A
frequency communication services. The user
might need to take mitigation measures, such
as relocating or re-orienting the equipment.
Fulfilled
Test level of profes-
sional medical
equipment and ba-
sic EMC standards
+ 8 kV contact dis-
charge
+ 15 kV air dis-
charge
+ 2 kV mains
+ 1 kV data lines
+ 1 kV line-line volt-
age
+ 2 kV line-ground
voltage
• 0% U
for ½ peri-
r
od
• 0% U
for 1 peri-
r
od
• 70% U
(30%
r
breakdown of
Electromagnetic Environ-
ment Guidelines
Floors should consist of
wood, concrete or ceramic
tiles. The relative humidity
must be at least 30%, if the
floor is made of synthetic
material.
The quality of the voltage
supplied should correspond
to a typical commercial or
clinical environment.
The quality of the voltage
supplied should correspond
to that of a typical commer-
cial or clinical environment.
The quality of the voltage
supply should correspond to
that of a typical commercial
or clinical environment.