Useful Information - IKA HABITAT Manual

Hide thumbs Also See for HABITAT:
Table of Contents

Advertisement

3
1
5
6
10
HA.cab.sw
1
Filter
2
Filter heater
3
Plug in barb connector 6 mm x 5 mm
4
Hose reducer 6.4 mm x 3.9 mm
5
Y hose connector 3.2 mm
6
Silicone tube 3.2 mm x 6.4 mm x 15 m
7
Silicone tube 4.8 mm x 9.5 mm x 3 m
8
pH cable
12
HA.cab.sw
2
3
8
7
8
11
12
HA.cab.dw
15
14
9
DO cable
10
USB cable
11
RS 232 cable
12
Tubing clamp
13
Silicone cap
14
Hose with GL18 and quick coupling
15
Silicone tube 8 mm x 12.7 mm x 5 m
16
Hose clamp
4
4
9
1
Single-use vessel
13
2
HA.s.tm.2 (Temperature sensor)
3
HA.s.do.2 (DO sensor)

Useful information

› The HABITAT is operated by an Android or
16
iOs tablet PC, which is part of the delivery
scope.
› As an alternative the HABITAT Graphic-User
Interface (GUI) software is also available for
Windows based tablet PC / notebooks.
› Wired and wireless operation is supported.
Monitoring is also possible by using the Bi-
oreactor App for Android and iOs mobile
systems.
› During operation the tablet PC / notebook
can be removed from the device. The con-
trol tower is working independently from
the GUI software. All data is saved in the
control tower.
› The main parameter status is shown by an
LED light bar.
Green light: ok, within the range
Red light: not ok, out of the range
› If the tablet PC / notebook is connected
again, all data will be transferred from the
control tower to the GUI to be displayed.
› Safe-STOP button: To immediately pause
all actuators device (Stirrer, Heater, Pumps,
MFCs, Thermostat, Light)
Safe-STOP: Press the button at the control
tower / or the button on GUI. All actuators
stop.
Exit Safe-STOP: Press the button at the con-
trol tower only! All actuators resume back to
1
2
3
4
5
HA.cab.sw
4
SU holder
5
HA.cab.sw
6
HA.sf.250 (Bottle 250 ml x4)
the setting before mode entry.
› Mains switch: Power-ON / Power-OFF the
device.
Power-ON: Control tower on, ready to use.
Power-OFF: Control tower off, no power.
!
Notice!
› Data log file is created only when starting the
"Experiment", and it records the data all the
time, from the start to the stop of the Exper-
iment, by pressing the PLAY button on the
user interface software. Only one filename is
created for one experiment (start-stop).
› In safe-STOP mode, the data log continues
recording. Log file will show all actuators
that have been stopped during the period of
time.
Single-use vessel
6
safe
STOP
!
Mains switch
13

Advertisement

Table of Contents
loading

Table of Contents