IKA HABITAT Manual page 27

Hide thumbs Also See for HABITAT:
Table of Contents

Advertisement

2.5.8 Foam limit Control
Mode:
› Off: The sensor will detect the foam level but not take any action.
› Auto: If foam is detected the controller will control the Anti-foam pump to fill in Anti-foam.
On delay: The time delay after the foam is detected (Activate status) before the Anti-foam fill-in.
On time: Pump duration on time.
Sensitivity: To adjust the sensitivity of the foam detection sensor.
Status Indicator: To indicate the status of the sensor (Normal/Activate).
2.5.9 Level limit Control
Mode:
› Off: The sensor will detect the level but not take any action.
› Auto: If the level is active the controller will control the Level pump to feed / harvest.
On delay: The time delay after the level sensor is active before the Level pump feed / harvest.
On time: Pump duration on time.
Sensitivity: To adjust the sensitivity of the level detection sensor.
Status Indicator: To indicate status of the sensor (Normal / Activate).
Level control mode: To control the direction of the pump: Harvest = suck out / Feed = Fill-in.
52
2.5.10 Temperature Control
Mode:
› Off: Turn off
› Auto – Control: Vessel temperature is following the user set point.
› Profile: To set pre-defined group of set value (up to 10) which will be executed in a period of
pre-defined time.
Set value offset: To set the temperature offset from the set point.
Actuator: Select Heating blanket / Thermostat (HRC 2) to control the vessel temperature.
2.5.11 pH Control
Mode:
› Off: Turn off
› Auto: Control process pH value is following user set point (feedback control).
› Profile: To set pre-defined group of set value (up to 10) which will be executed in a period of
pre-defined time.
Set value offset: To set the pH offset from the set point.
53