4-3-6. Balance weight

Preset the value of the load (weight) to be applied to the load cell during span calibration.
Setting range / 00000 to 99999
CC-Link
MSB
H
L
Console
1) Select code 40.
2) Press the ENT key, and then set the bal-
ance weight.
3) Determine the setting with the ENT key.
Write the code No. and set value.
0
Code No.
00000 to 99999
Set value
Initial value / 10000
Word size
remote register
LSB
RWw0006
40
RWw0005
RWw0004
Balance weight
(up to 5 digits)